Specifications
Table Of Contents
- E6392-90042.pdf
- E6392-90042_manual
- Front Matter
- Contents
- 1 Preparing for Use
- 2 Programming Command Guidelines
- 3 Programming Command Reference
- IEEE Common Command
- CONFigure Subsystem
- DISPlay Subsystem
- HCOPy Subsystem
- RFGenerator Subsystem
- SMONitor Subsystem
- SYSTem Subsystem
- TESTs Subsystem for AUTOMATIC TEST
- All Tests Results
- All Tests Results
- Command Reference for Burst Timing/Power Ramp Measurement Screen
- Command Reference for DC Current Measurement Screen
- Command Reference for MS Information
- Command Reference for Peak TX Power Measurement Screen
- Command Reference for Phase/Frequency Error Measurement Screen
- Command Reference for Sensitivity/RX Quality/RX Level Measurement Screen
- Command Reference for Stand-by/Measuring Screens
- TESTs Subsystem for MANUAL TEST
- Command Reference for DC Current Measurement Screen
- Command Reference for Measuring Screens
- Command Reference for MS Information
- Command Reference for Peak TX Power /Burst Timing/Power Ramp Measurement Screen
- Command Reference for Phase/Frequency Error Measurement Screen
- Command Reference for Sensitivity/RX Quality/RX Level Measurement Screen
- Command Reference for Spectrum Monitor Measurement Screen
- Command Reference for Stand-by Screens
- TRIGger Subsystem
- 4 Programming Command Cross Reference
- 5 Example Programs
- A Syntax Diagram
- B Command Difference between E6392A & E6392B
- Index

Chapter 3 87
Programming Command Reference
TESTs Subsystem for MANUAL TEST
3. Programming Command
Reference
This command specifies the frequency of Traffic Channel for each of the radio
standards in conjunction with the Multi Band settings. The allowable range and
preset values are as follows:
Command Reference for Measuring Screens
These commands are used to set the controls and parameters associated with an
initial test condition.
Send the "DISPlay:MANual:MEASure" command to display the Measuring for
MANUAL TEST before sending the following commands.
BS Call
TESTs:MANual:MEASure:BSCall
This command executes a call setup to the mobile from the Test Set. There is no
query form of this command.
NOTE If Option 002 and Option 040 are installed and the mode is set to
ASYNC or GPRS, this command is invalid.
BS Level
TESTs:MANual:MEASure:AMPLitude <real> [dBm]
TESTs:MANual:MEASure:AMPLitude?
This command specifies an amplitude of BS Level in the Measuring screen. The
allowable range is from −110.0 to −50.0 (dBm) in 0.5 steps. At preset, this is set to
−50.0 (dBm).
RFORmat MBANd Allowable Range [Hz] Preset Value [Hz]
GSM900
DCS1800
890200000 to 914800000;
1710200000 to 1784800000
896000000
OFF 890200000 to 914800000
E-GSM
DCS1800
880200000 to 914800000;
1710200000 to 1784800000
896000000
OFF 880200000 to 914800000
DCS1800
GSM900
1710200000 to 1784800000;
890200000 to 914800000
1710200000
E-GSM
1710200000 to 1784800000;
880200000 to 914800000
OFF 1710200000 to 1784800000
PCS1900 OFF 1850200000 to 1909800000 1850200000