84 Chapter 3
Programming Command Reference
TESTs Subsystem for MANUAL TEST
TESTs Subsystem for MANUAL TEST
This function allows you to execute real time measurements. Depending on the
combination of the options installed with the Test Set, the different measurement
modes are available as follows:
NOTE For Test Sets with Option 002, the DC Power voltage setting done
in this function are copied to the AUTOMATIC TEST, SIGNAL
GENERATOR and SPECTRUM MONITOR and vice versa.
Command Reference for DC Current Measurement Screen
The Current commands are used to set the controls and parameters associated with
the DC Current measurements and the pass/fail tests.
Send the "DISPlay:MANual:CURRent" command to display the DC Current
measurement screen for MANUAL TEST before sending the following
commands.
DC Current Measurement (ASYNC or GPRS mode)
TESTs:MANual:CURRent:RESults?
This query command returns a set of test result values, <string 1>,
<string 2> and <string 3>, for asynchronous or GPRS mode tests of DC
Current (without titling Camp On nor Talk).
<string 1> (1 digit) is 0 when all values are within the normal range, or 1 if any
measurement is overranged, underranged or not made.
<string 2> (1 digit) is the binary data, 0 or 1, for the pass (0) or fail (1) result
using the low and high limits for Talk.
<string 3> (5 digits) is the measured dc current data ranging from 0.000 to
1.000 (A). If any measurement is overranged, underranged or not made, the
value 2.500 is returned.
NOTE This command is valid only for Test Sets with Option 002 and
Option 040.
Option Available Measurement Mode
<none> SYNC (Synchronous)
002 SYNC, ASYNC (Asynchronous)
040 SYNC, GPRS
002, 040 SYNC, ASYNC, GPRS