Specifications
Table Of Contents
- E6392-90042.pdf
- E6392-90042_manual
- Front Matter
- Contents
- 1 Preparing for Use
- 2 Programming Command Guidelines
- 3 Programming Command Reference
- IEEE Common Command
- CONFigure Subsystem
- DISPlay Subsystem
- HCOPy Subsystem
- RFGenerator Subsystem
- SMONitor Subsystem
- SYSTem Subsystem
- TESTs Subsystem for AUTOMATIC TEST
- All Tests Results
- All Tests Results
- Command Reference for Burst Timing/Power Ramp Measurement Screen
- Command Reference for DC Current Measurement Screen
- Command Reference for MS Information
- Command Reference for Peak TX Power Measurement Screen
- Command Reference for Phase/Frequency Error Measurement Screen
- Command Reference for Sensitivity/RX Quality/RX Level Measurement Screen
- Command Reference for Stand-by/Measuring Screens
- TESTs Subsystem for MANUAL TEST
- Command Reference for DC Current Measurement Screen
- Command Reference for Measuring Screens
- Command Reference for MS Information
- Command Reference for Peak TX Power /Burst Timing/Power Ramp Measurement Screen
- Command Reference for Phase/Frequency Error Measurement Screen
- Command Reference for Sensitivity/RX Quality/RX Level Measurement Screen
- Command Reference for Spectrum Monitor Measurement Screen
- Command Reference for Stand-by Screens
- TRIGger Subsystem
- 4 Programming Command Cross Reference
- 5 Example Programs
- A Syntax Diagram
- B Command Difference between E6392A & E6392B
- Index

Chapter 3 77
Programming Command Reference
TESTs Subsystem for AUTOMATIC TEST
3. Programming Command
Reference
• <string 2> (1 digit) is the binary data, 0 or 1, for the pass (0) or fail (1) result.
• <string 3> (6 digits) is the marker position ranging from 000.00 to 147.00 in
bits.
• <string 4> (5 digits) is the measured error ranging from −99.9 to +99.9 in °
(degree). If its measurement is overranged, underranged or not made, the value
+99.9 is returned.
Phase Error Marker Position Control
TESTs:AUTO:RESults:PHASe1:MARKer:POINt UP|DOWM
TESTs:AUTO:RESults:PHASe1:MARKer:POINt <real>
TESTs:AUTO:RESults:PHASe1:MARKer:POINt?
TESTs:AUTO:RESults:PHASe2:MARKer:POINt UP|DOWN
TESTs:AUTO:RESults:PHASe2:MARKer:POINt <real>
TESTs:AUTO:RESults:PHASe2:MARKer:POINt?
TESTs:AUTO:RESults:PHASe3:MARKer:POINt UP|DOWN
TESTs:AUTO:RESults:PHASe3:MARKer:POINt <real>
TESTs:AUTO:RESults:PHASe3:MARKer:POINt?
TESTs:AUTO:RESults:PHASe4:MARKer:POINt UP|DOWM
TESTs:AUTO:RESults:PHASe4:MARKer:POINt <real>
TESTs:AUTO:RESults:PHASe4:MARKer:POINt?
TESTs:AUTO:RESults:PHASe5:MARKer:POINt UP|DOWN
TESTs:AUTO:RESults:PHASe5:MARKer:POINt <real>
TESTs:AUTO:RESults:PHASe5:MARKer:POINt?
TESTs:AUTO:RESults:PHASe6:MARKer:POINt UP|DOWN
TESTs:AUTO:RESults:PHASe6:MARKer:POINt <real>
TESTs:AUTO:RESults:PHASe6:MARKer:POINt?
This command enables you to move the marker UP or DOWN in steps, or by
<real> bits ranging from 0.0 to 147.0 bits, in the Phase Error graph for each of the
six measurements. At preset, these are set to 0.25 (bits).
Phase Error Measurement
TESTs:AUTO:RESults:PHASe1?
TESTs:AUTO:RESults:PHASe2?
TESTs:AUTO:RESults:PHASe3?
TESTs:AUTO:RESults:PHASe4?
TESTs:AUTO:RESults:PHASe5?
TESTs:AUTO:RESults:PHASe6?
This query command returns a set of test result values, <string 1>,
<string 2>, <string 3>, and <string 4>, for each of the six
measurements of Peak and RMS Phase Error.
• <string 1> (1 digit) is 0 when all values are within the normal range, or 1 if any
measurement is overranged, underranged or not made.