Specifications
Table Of Contents
- E6392-90042.pdf
- E6392-90042_manual
- Front Matter
- Contents
- 1 Preparing for Use
- 2 Programming Command Guidelines
- 3 Programming Command Reference
- IEEE Common Command
- CONFigure Subsystem
- DISPlay Subsystem
- HCOPy Subsystem
- RFGenerator Subsystem
- SMONitor Subsystem
- SYSTem Subsystem
- TESTs Subsystem for AUTOMATIC TEST
- All Tests Results
- All Tests Results
- Command Reference for Burst Timing/Power Ramp Measurement Screen
- Command Reference for DC Current Measurement Screen
- Command Reference for MS Information
- Command Reference for Peak TX Power Measurement Screen
- Command Reference for Phase/Frequency Error Measurement Screen
- Command Reference for Sensitivity/RX Quality/RX Level Measurement Screen
- Command Reference for Stand-by/Measuring Screens
- TESTs Subsystem for MANUAL TEST
- Command Reference for DC Current Measurement Screen
- Command Reference for Measuring Screens
- Command Reference for MS Information
- Command Reference for Peak TX Power /Burst Timing/Power Ramp Measurement Screen
- Command Reference for Phase/Frequency Error Measurement Screen
- Command Reference for Sensitivity/RX Quality/RX Level Measurement Screen
- Command Reference for Spectrum Monitor Measurement Screen
- Command Reference for Stand-by Screens
- TRIGger Subsystem
- 4 Programming Command Cross Reference
- 5 Example Programs
- A Syntax Diagram
- B Command Difference between E6392A & E6392B
- Index

Chapter 3 75
Programming Command Reference
TESTs Subsystem for AUTOMATIC TEST
3. Programming Command
Reference
The choices for each bit are 1 for Fail and 0 for Pass.
• <string 3> (5 digits) and <string 4> (5 digits) are the measured data ranging
from 0.000 to 1.000 (A) at Camp On and Talk. If any measurement is
overranged, underranged or not made, the value 2.5 (A) is returned.
Command Reference for MS Information
MS Information
TESTs:AUTO:RESults:MSINformation:IMSI?
This query command returns the IMSI string up to 15 letters.
TESTs:AUTO:RESults:MSINformation:IMEI?
This query command returns the IMEI string up to 15 letters.
TESTs:AUTO:RESults:MSINformation:DIAL?
This query command returns the dialed number string up to 20 digits.
TESTs:AUTO:RESults:MSINformation:PCLass?
This query command returns a 1-digit power class string value.
TESTs:AUTO:RESults:MSINformation:GVERsion?
This query command returns the 1-digit version number of GSM mobile under
test.
Command Reference for Peak TX Power Measurement Screen
Send the "DISPlay:AUTO:TPOWer" command to display the Peak TX Power
measurement screen for AUTOMATIC TEST before sending the following
commands.
Peak TX Power Measurement
TESTs:AUTO:RESults:TPOWer1?
TESTs:AUTO:RESults:TPOWer2?
TESTs:AUTO:RESults:TPOWer3?
TESTs:AUTO:RESults:TPOWer4?
TESTs:AUTO:RESults:TPOWer5?
TESTs:AUTO:RESults:TPOWer6?
These query commands return a set of test result values, <string 1>,
<string 2>, <string 3>, <string 4>, and <string 5>, for each of
the six measurements of Peak TX Power.
• <string 1> (1 digit) is 0 when all values are within the normal range, or 1 if any
measurement is overranged, underranged or not made.
• <string 2> (1 digit) is the binary coded decimal ranging from 0 to 7 for the
pass/fail results as follows: