Specifications

Table Of Contents
64 Chapter 3
Programming Command Reference
SMONitor Subsystem
SMONitor Subsystem
The SMONitor subsystem commands are used to set and read the controls and
parameters of the SPECTRUM MONITOR function. Send the
"DISPlay:SMONitor" command to display the SPECTRUM MONITOR screen
before sending the following commands.
NOTE This section is applicable only for Test Sets with Option 002.
NOTE For Test Sets with Option 002, the Radio Standard and Channel
settings done in this function are copied to SPECTRUM
MONITOR and vice versa.
SMONitor Subsystem Command Reference
Averaging Function
SMONitor:AVERage OFF|<int>
SMONitor:AVERage?
This command defines whether or not to activate the averaging function. The
choices are OFF and <int> for averaging time. The allowable range is from 2 to 99.
At preset, this is set to OFF.
DC Power Mode
SMONitor:DCPower:MODE ON|OFF|1|0
SMONitor:DCPower:MODE?
This command defines a supply mode of DC Power. The choices are ON (1) and
OFF (0). If set to ON (1), DC Power is always supplied. If set to OFF (0), DC
Power is not supplied. At preset, this is set to OFF.
DC Power Voltage
SMONitor:DCPower:VOLT <real> [V]
SMONitor:DCPower:VOLT?
This command specifies a voltage value of DC Power. The allowable range is from
3.0 to 11.0 (V) in 0.1 steps. At preset, this is set to 3.0 (V).
Display Span
SMONitor:SPAN 400k|100k
SMONitor:SPAN?
This command specifies a frequency bandwidth of Span. The choices are 400k for
the 0 to 400 kHz span, and 100k for the ±100 kHz span relative to the Traffic
Channel frequency. At preset, this is set to 400k.