Specifications
Table Of Contents
- E6392-90042.pdf
- E6392-90042_manual
- Front Matter
- Contents
- 1 Preparing for Use
- 2 Programming Command Guidelines
- 3 Programming Command Reference
- IEEE Common Command
- CONFigure Subsystem
- DISPlay Subsystem
- HCOPy Subsystem
- RFGenerator Subsystem
- SMONitor Subsystem
- SYSTem Subsystem
- TESTs Subsystem for AUTOMATIC TEST
- All Tests Results
- All Tests Results
- Command Reference for Burst Timing/Power Ramp Measurement Screen
- Command Reference for DC Current Measurement Screen
- Command Reference for MS Information
- Command Reference for Peak TX Power Measurement Screen
- Command Reference for Phase/Frequency Error Measurement Screen
- Command Reference for Sensitivity/RX Quality/RX Level Measurement Screen
- Command Reference for Stand-by/Measuring Screens
- TESTs Subsystem for MANUAL TEST
- Command Reference for DC Current Measurement Screen
- Command Reference for Measuring Screens
- Command Reference for MS Information
- Command Reference for Peak TX Power /Burst Timing/Power Ramp Measurement Screen
- Command Reference for Phase/Frequency Error Measurement Screen
- Command Reference for Sensitivity/RX Quality/RX Level Measurement Screen
- Command Reference for Spectrum Monitor Measurement Screen
- Command Reference for Stand-by Screens
- TRIGger Subsystem
- 4 Programming Command Cross Reference
- 5 Example Programs
- A Syntax Diagram
- B Command Difference between E6392A & E6392B
- Index

Chapter 3 61
Programming Command Reference
RFGenerator Subsystem
3. Programming Command
Reference
RFGenerator Subsystem
The RFGenerator subsystem commands are used to set and read the controls and
parameters of the SIGNAL GENERATOR functions. Send the
"DISPlay:RFGenerator" command to display the SIGNAL GENERATOR screen
before sending the following commands.
NOTE This section is applicable only for Test Sets with Option 002.
NOTE For Test Sets with Option 002, the Radio Standard and Channel
settings done in this function are copied to SPECTRUM
MONITOR and vice versa.
RFGenerator Subsystem Command Reference
Amplitude
RFGenerator:AMPLitude <real> [DB]
RFGenerator:AMPLitude?
This command specifies an output amplitude of RF Level. The allowable range is
from −110.0 to −50.0 (dBm) in 0.5 steps. At preset, this is set to −50.0 (dBm).
DC Power Mode
RFGenerator:DCPower:MODE ON|OFF|1|0
RFGenerator:DCPower:MODE?
This command defines a supply mode of DC Power. The choices are ON (1) and
OFF (0). If set to ON (1), DC Power is always supplied. If set to OFF (0), DC
Power is not supplied. At preset, this is set to OFF.
DC Power Voltage
RFGenerator:DCPower:VOLT <real> [V]
RFGenerator:DCPower:VOLT?
This command specifies a voltage value of DC Power. The allowable range is from
3.0 to 11.0 (V) in 0.1 steps. At preset, this is set to 3.0 (V).
Modulation Function
RFGenerator:MODulation OFF|PN9|ALL0
RFGenerator:MODulation?
This command defines a type of Modulation. The choices are OFF, PN9 and
ALL_0. At preset, this is set to OFF.
• If OFF is selected, the RF signal is a continuous wave.