Specifications
Table Of Contents
- E6392-90042.pdf
- E6392-90042_manual
- Front Matter
- Contents
- 1 Preparing for Use
- 2 Programming Command Guidelines
- 3 Programming Command Reference
- IEEE Common Command
- CONFigure Subsystem
- DISPlay Subsystem
- HCOPy Subsystem
- RFGenerator Subsystem
- SMONitor Subsystem
- SYSTem Subsystem
- TESTs Subsystem for AUTOMATIC TEST
- All Tests Results
- All Tests Results
- Command Reference for Burst Timing/Power Ramp Measurement Screen
- Command Reference for DC Current Measurement Screen
- Command Reference for MS Information
- Command Reference for Peak TX Power Measurement Screen
- Command Reference for Phase/Frequency Error Measurement Screen
- Command Reference for Sensitivity/RX Quality/RX Level Measurement Screen
- Command Reference for Stand-by/Measuring Screens
- TESTs Subsystem for MANUAL TEST
- Command Reference for DC Current Measurement Screen
- Command Reference for Measuring Screens
- Command Reference for MS Information
- Command Reference for Peak TX Power /Burst Timing/Power Ramp Measurement Screen
- Command Reference for Phase/Frequency Error Measurement Screen
- Command Reference for Sensitivity/RX Quality/RX Level Measurement Screen
- Command Reference for Spectrum Monitor Measurement Screen
- Command Reference for Stand-by Screens
- TRIGger Subsystem
- 4 Programming Command Cross Reference
- 5 Example Programs
- A Syntax Diagram
- B Command Difference between E6392A & E6392B
- Index

52 Chapter 3
Programming Command Reference
CONFigure Subsystem
Phase Error Test
CONFigure:SEQuence:PHASe <int>
CONFigure:SEQuence:PHASe?
This command defines whether or not to include Phase Error tests in the test
sequence for the six measurements. <int> is a binary coded decimal ranging from 0
to 63 as follows:
bit 0 (1): For testing TCH1 or DCP1
bit 1 (2): For testing TCH2 or DCP2
bit 2 (4): For testing TCH3 or DCP3
bit 3 (8): For testing TCH4 or DCP4
bit 4 (16): For testing TCH5 or DCP5
bit 5 (32): For testing TCH6 or DCP6
The choices for each bit are 1 for Run and 0 for --- (skip). At preset, this is set to
63.
Power Ramp Test
CONFigure:SEQuence:RAMP <int>
CONFigure:SEQuence:RAMP?
This command defines whether or not to include Power Ramp tests in the test
sequence for the six measurements. <int> is a binary coded decimal ranging from 0
to 63 as follows:
bit 0 (1): For testing TCH1 or DCP1
bit 1 (2): For testing TCH2 or DCP2
bit 2 (4): For testing TCH3 or DCP3
bit 3 (8): For testing TCH4 or DCP4
bit 4 (16): For testing TCH5 or DCP5
bit 5 (32): For testing TCH6 or DCP6
The choices for each bit are 1 for Run and 0 for --- (skip). At preset, this is set to
63.
Radio Standard
CONFigure:SEQuence:RFORmat GSM900|E-GSM|DCS1800|PCS1900
CONFigure:SEQuence:RFORmat?
This command defines a system of Radio Standard to be tested. The choices are
GSM900, E-GSM, DCS1800, and PCS1900. At preset, this is set to GSM900. If
PCS1900 is selected, the Multi Band command is automatically set to OFF.