Specifications
Table Of Contents
- E6392-90042.pdf
- E6392-90042_manual
- Front Matter
- Contents
- 1 Preparing for Use
- 2 Programming Command Guidelines
- 3 Programming Command Reference
- IEEE Common Command
- CONFigure Subsystem
- DISPlay Subsystem
- HCOPy Subsystem
- RFGenerator Subsystem
- SMONitor Subsystem
- SYSTem Subsystem
- TESTs Subsystem for AUTOMATIC TEST
- All Tests Results
- All Tests Results
- Command Reference for Burst Timing/Power Ramp Measurement Screen
- Command Reference for DC Current Measurement Screen
- Command Reference for MS Information
- Command Reference for Peak TX Power Measurement Screen
- Command Reference for Phase/Frequency Error Measurement Screen
- Command Reference for Sensitivity/RX Quality/RX Level Measurement Screen
- Command Reference for Stand-by/Measuring Screens
- TESTs Subsystem for MANUAL TEST
- Command Reference for DC Current Measurement Screen
- Command Reference for Measuring Screens
- Command Reference for MS Information
- Command Reference for Peak TX Power /Burst Timing/Power Ramp Measurement Screen
- Command Reference for Phase/Frequency Error Measurement Screen
- Command Reference for Sensitivity/RX Quality/RX Level Measurement Screen
- Command Reference for Spectrum Monitor Measurement Screen
- Command Reference for Stand-by Screens
- TRIGger Subsystem
- 4 Programming Command Cross Reference
- 5 Example Programs
- A Syntax Diagram
- B Command Difference between E6392A & E6392B
- Index

Chapter 3 49
Programming Command Reference
CONFigure Subsystem
3. Programming Command
Reference
The choices for each bit are 1 for Run and 0 for --- (skip). At preset, this is set to
63.
DC Current Test
CONFigure:SEQuence:CURRent <int>
CONFigure:SEQuence:CURRent?
This command defines whether or not to include DC Current tests in the test
sequence for the six measurements. <int> is a binary coded decimal ranging from 0
to 63 as follows:
bit 0 (1): For testing TCH1 or DCP1
bit 1 (2): For testing TCH2 or DCP2
bit 2 (4): For testing TCH3 or DCP3
bit 3 (8): For testing TCH4 or DCP4
bit 4 (16): For testing TCH5 or DCP5
bit 5 (32): For testing TCH6 or DCP6
The choices for each bit are 1 for Run and 0 for --- (skip). At preset, this is set to
63.
DC Power Mode
CONFigure:SEQuence:DCPower:MODE AUTO|ON|OFF
CONFigure:SEQuence:DCPower:MODE?
This command defines a supply mode of DC Power. The choices are AUTO, ON,
and OFF. At preset, this is set to OFF.
• If set to AUTO, DC Power is supplied during a measurement cycle or until a
measurement is aborted.
• If set to ON, DC Power is always supplied.
• If set to OFF, DC Power is not supplied and the Variable field is automatically
set to TCH.
DC Power Voltage
CONFigure:SEQuence:DCPower:VOLT <real> [V]
CONFigure:SEQuence:DCPower:VOLT?
This command specifies a voltage value of DC Power. The allowable range is from
3.0 to 11.0 (V) in 0.1 steps. At preset, this is set to 3.0 (V).
Frequency Error Test
CONFigure:SEQuence:FREQuency <int>
CONFigure:SEQuence:FREQuency?