The choices for each bit are 1 for Run and 0 for --- (skip). At preset, this is set to
63.
DC Current Test
CONFigure:SEQuence:CURRent <int>
CONFigure:SEQuence:CURRent?
This command defines whether or not to include DC Current tests in the test
sequence for the six measurements. <int> is a binary coded decimal ranging from 0
to 63 as follows:
bit 0 (1): For testing TCH1 or DCP1
bit 1 (2): For testing TCH2 or DCP2
bit 2 (4): For testing TCH3 or DCP3
bit 3 (8): For testing TCH4 or DCP4
bit 4 (16): For testing TCH5 or DCP5
bit 5 (32): For testing TCH6 or DCP6
The choices for each bit are 1 for Run and 0 for --- (skip). At preset, this is set to
63.
DC Power Mode
CONFigure:SEQuence:DCPower:MODE AUTO|ON|OFF
CONFigure:SEQuence:DCPower:MODE?
This command defines a supply mode of DC Power. The choices are AUTO, ON,
and OFF. At preset, this is set to OFF.
If set to AUTO, DC Power is supplied during a measurement cycle or until a
measurement is aborted.
If set to ON, DC Power is always supplied.
If set to OFF, DC Power is not supplied and the Variable field is automatically
set to TCH.
DC Power Voltage
CONFigure:SEQuence:DCPower:VOLT <real> [V]
CONFigure:SEQuence:DCPower:VOLT?
This command specifies a voltage value of DC Power. The allowable range is from
3.0 to 11.0 (V) in 0.1 steps. At preset, this is set to 3.0 (V).
Frequency Error Test
CONFigure:SEQuence:FREQuency <int>
CONFigure:SEQuence:FREQuency?