Specifications
Table Of Contents
- E6392-90042.pdf
- E6392-90042_manual
- Front Matter
- Contents
- 1 Preparing for Use
- 2 Programming Command Guidelines
- 3 Programming Command Reference
- IEEE Common Command
- CONFigure Subsystem
- DISPlay Subsystem
- HCOPy Subsystem
- RFGenerator Subsystem
- SMONitor Subsystem
- SYSTem Subsystem
- TESTs Subsystem for AUTOMATIC TEST
- All Tests Results
- All Tests Results
- Command Reference for Burst Timing/Power Ramp Measurement Screen
- Command Reference for DC Current Measurement Screen
- Command Reference for MS Information
- Command Reference for Peak TX Power Measurement Screen
- Command Reference for Phase/Frequency Error Measurement Screen
- Command Reference for Sensitivity/RX Quality/RX Level Measurement Screen
- Command Reference for Stand-by/Measuring Screens
- TESTs Subsystem for MANUAL TEST
- Command Reference for DC Current Measurement Screen
- Command Reference for Measuring Screens
- Command Reference for MS Information
- Command Reference for Peak TX Power /Burst Timing/Power Ramp Measurement Screen
- Command Reference for Phase/Frequency Error Measurement Screen
- Command Reference for Sensitivity/RX Quality/RX Level Measurement Screen
- Command Reference for Spectrum Monitor Measurement Screen
- Command Reference for Stand-by Screens
- TRIGger Subsystem
- 4 Programming Command Cross Reference
- 5 Example Programs
- A Syntax Diagram
- B Command Difference between E6392A & E6392B
- Index

Chapter 3 43
Programming Command Reference
CONFigure Subsystem
3. Programming Command
Reference
Peak TX Power Limit
CONFigure:CONDition:LIMit:TPOWer
<int>,<real l>,<real h> [DB]
CONFigure:CONDition:LIMit:TPOWer?
This command specifies a value of Power Control Level (PWR CNTL) and the
low and high limits for Peak TX Power tests, for each radio standard set by the
Radio Format command. The allowable limit range for <real l> and <real h> is
from −99.9 to +99.9 (dB) in 0.1 steps. At preset, these are set to −3.0 and +3.0
(dB); an exception is that, for LOW level tests of DCS1800 or PCS1900, the preset
values are −4.0 and +4.0 (dB). The power control level and the preset values are as
follows:
Phase Error Limit
CONFigure:CONDition:LIMit:PHASe:PEAK <real h> [DEG]
CONFigure:CONDition:LIMit:PHASe:PEAK?
CONFigure:CONDition:LIMit:PHASe:RMS <real h> [DEG]
CONFigure:CONDition:LIMit:PHASe:RMS?
This command specifies the high limit for Peak or RMS Phase Error tests. The
allowable range is from 0.0 to 99.9 (°) in 0.1 steps. At preset, these are set to 20.0
(°) for peak and 5.0 (°) for rms.
Power Control Level
CONFigure:CONDition:PLEVel:HIGH <int>
CONFigure:CONDition:PLEVel:HIGH?
CONFigure:CONDition:PLEVel:MEDium <int>
CONFigure:CONDition:PLEVel:MEDium?
CONFigure:CONDition:PLEVel:LOW <int>
CONFigure:CONDition:PLEVel:LOW?
CONFigure:CONDition:PLEVel:MANual <int>
CONFigure:CONDition:PLEVel:MANual?
RFORmat Allowable Range for <int> Preset Value
GSM900 2 (+39 dBm) to 19 (+5 dBm) 5 (+33 dBm)
E-GSM 2 (+39 dBm) to 19 (+5 dBm) 5 (+33 dBm)
DCS1800 0 (+30 dBm) to 15 (+0 dBm), or
29 (+36 dBm) to 31 (+32 dBm)
0 (+30 dBm)
PCS1900 0 (+30 dBm) to 15 (+0 dBm), or
30 (+33 dBm) to 31 (+32 dBm)
0 (+30 dBm)