Specifications
Table Of Contents
- E6392-90042.pdf
- E6392-90042_manual
- Front Matter
- Contents
- 1 Preparing for Use
- 2 Programming Command Guidelines
- 3 Programming Command Reference
- IEEE Common Command
- CONFigure Subsystem
- DISPlay Subsystem
- HCOPy Subsystem
- RFGenerator Subsystem
- SMONitor Subsystem
- SYSTem Subsystem
- TESTs Subsystem for AUTOMATIC TEST
- All Tests Results
- All Tests Results
- Command Reference for Burst Timing/Power Ramp Measurement Screen
- Command Reference for DC Current Measurement Screen
- Command Reference for MS Information
- Command Reference for Peak TX Power Measurement Screen
- Command Reference for Phase/Frequency Error Measurement Screen
- Command Reference for Sensitivity/RX Quality/RX Level Measurement Screen
- Command Reference for Stand-by/Measuring Screens
- TESTs Subsystem for MANUAL TEST
- Command Reference for DC Current Measurement Screen
- Command Reference for Measuring Screens
- Command Reference for MS Information
- Command Reference for Peak TX Power /Burst Timing/Power Ramp Measurement Screen
- Command Reference for Phase/Frequency Error Measurement Screen
- Command Reference for Sensitivity/RX Quality/RX Level Measurement Screen
- Command Reference for Spectrum Monitor Measurement Screen
- Command Reference for Stand-by Screens
- TRIGger Subsystem
- 4 Programming Command Cross Reference
- 5 Example Programs
- A Syntax Diagram
- B Command Difference between E6392A & E6392B
- Index

Chapter 3 41
Programming Command Reference
CONFigure Subsystem
3. Programming Command
Reference
NOTE If Option 040 is installed and the mode is set to GPRS, BER can
noe is measured in MANUAL TEST.
BLER Limit
CONFigure:CONDition:LIMit:BLER <real h> [PCT]
CONFigure:CONDition:LIMit:BLER?
This command specifies the high limit for BLER (block error ratio) tests if Option
040 is installed and the GPRS mode is selected. The allowable range is from 0.00
to 99.99 (%) in 0.01 steps. At preset, this is set to 2.44 (%).
NOTE If the mode is set to GPRS on the Test Set with Option 040,
Sensitivity in BLER is measured in MANUAL TEST.
BS Level
CONFigure:CONDition:AMPLitude <real> [DBM]
CONFigure:CONDition:AMPLitude?
This command specifies an amplitude of the Test Set applied to the mobile under
test. The allowable range is from −110.0 to −50.0 (dBm) in 0.5 steps. At preset,
this is set to −50.0 (dBm).
Burst Timing Limit
CONFigure:CONDition:LIMit:BTIMing
<real l>,<real h> [Bits]
CONFigure:CONDition:LIMit:BTIMing?
This command specifies the low and high limits for Burst Timing tests. The
allowable range is from −9.9 to +9.9 (bits) in 0.1 steps. At preset, these are set to
−1.0 and +1.0 (bits).
DC Current Limit
CONFigure:CONDition:LIMit:CURRent:CAMP
<int l>,<int h> [MA]
CONFigure:CONDition:LIMit:CURRent:CAMP?
CONFigure:CONDition:LIMit:CURRent:TALK
<int l>,<int h> [MA]
CONFigure:CONDition:LIMit:CURRent:TALK?
This command specifies the low and high limits for DC Current tests at Camp On
or Talk. The allowable range is from 3 to 1000 (mA) in 1 steps. At preset, these are
set to 50 and 200 (mA) for Camp On, and 300 and 700 (mA) for Talk, respectively.
NOTE If Option 002 is installed and the mode is set to ASYNC in
MANUAL TEST, the limits for Talk are used for DC Current tests.
If Option 040 is installed and the mode is set to GPRS in
MANUAL TEST, the limits for Talk are used for DC Current tests.