Specifications
Table Of Contents
- E6392-90042.pdf
- E6392-90042_manual
- Front Matter
- Contents
- 1 Preparing for Use
- 2 Programming Command Guidelines
- 3 Programming Command Reference
- IEEE Common Command
- CONFigure Subsystem
- DISPlay Subsystem
- HCOPy Subsystem
- RFGenerator Subsystem
- SMONitor Subsystem
- SYSTem Subsystem
- TESTs Subsystem for AUTOMATIC TEST
- All Tests Results
- All Tests Results
- Command Reference for Burst Timing/Power Ramp Measurement Screen
- Command Reference for DC Current Measurement Screen
- Command Reference for MS Information
- Command Reference for Peak TX Power Measurement Screen
- Command Reference for Phase/Frequency Error Measurement Screen
- Command Reference for Sensitivity/RX Quality/RX Level Measurement Screen
- Command Reference for Stand-by/Measuring Screens
- TESTs Subsystem for MANUAL TEST
- Command Reference for DC Current Measurement Screen
- Command Reference for Measuring Screens
- Command Reference for MS Information
- Command Reference for Peak TX Power /Burst Timing/Power Ramp Measurement Screen
- Command Reference for Phase/Frequency Error Measurement Screen
- Command Reference for Sensitivity/RX Quality/RX Level Measurement Screen
- Command Reference for Spectrum Monitor Measurement Screen
- Command Reference for Stand-by Screens
- TRIGger Subsystem
- 4 Programming Command Cross Reference
- 5 Example Programs
- A Syntax Diagram
- B Command Difference between E6392A & E6392B
- Index

40 Chapter 3
Programming Command Reference
CONFigure Subsystem
Command Reference for Test Setup: Test Condition Screen
The Condition commands are used to set the controls and parameters associated
with the test condition including the pass/fail test limits for AUTOMATIC TEST
and MANUAL TEST. The query commands return the settings of those controls
and parameters.
Send the "DISPlay:CONFigure:CONDition" command to display the Test Setup:
Test Condition screen before sending the following commands.
NOTE If Option 040 is installed and the mode is set to GPRS, the test
items FER, BER, RX Quality, and RX Level for sensitivity tests
can not be displayed in MANUAL TEST, however these can be
configured. One command for Sensitivity in BLER is added to the
GPRS mode.
Averaging Function
CONFigure:CONDition:AVERage OFF|<int>
CONFigure:CONDition:AVERage?
This command defines whether or not to activate the averaging function. The
choices are OFF and <int> for the number of averaging times. The allowable range
of <int> is from 2 to 99. At preset, this is set to OFF.
BER BS Level
CONFigure:CONDition:SENSitivity:LEVel <real> [DBM]
CONFigure:CONDition:SENSitivity:LEVel?
This command specifies an amplitude of the Test Set for error ratio tests, for each
radio standard set by the Radio Format command. The allowable range is from
−110.0 to −50.0 (dBm) in 0.5 steps. At preset, this is set to −102.0 (dBm) for
GSM900 and E-GSM, −100.0 (dBm) for DCS1800, or −102.0 (dBm) for
PCS1900.
BER Frames
CONFigure:CONDition:SENSitivity:FRAMe <int>
CONFigure:CONDition:SENSitivity:FRAMe?
This command specifies the number of the Test Set for error ratio tests. The
allowable range is from 1 to 13000 (frames) in 1 steps. At preset, this is set to 10
(frames).
BER Limit
CONFigure:CONDition:LIMit:BER <real h> [PCT]
CONFigure:CONDition:LIMit:BER?
This command specifies the high limit for BER (bit error ratio) tests. The
allowable range is from 0.00 to 99.99 (%) in 0.01 steps. At preset, this is set to 2.44
(%).