Specifications

Table Of Contents
40 Chapter 3
Programming Command Reference
CONFigure Subsystem
Command Reference for Test Setup: Test Condition Screen
The Condition commands are used to set the controls and parameters associated
with the test condition including the pass/fail test limits for AUTOMATIC TEST
and MANUAL TEST. The query commands return the settings of those controls
and parameters.
Send the "DISPlay:CONFigure:CONDition" command to display the Test Setup:
Test Condition screen before sending the following commands.
NOTE If Option 040 is installed and the mode is set to GPRS, the test
items FER, BER, RX Quality, and RX Level for sensitivity tests
can not be displayed in MANUAL TEST, however these can be
configured. One command for Sensitivity in BLER is added to the
GPRS mode.
Averaging Function
CONFigure:CONDition:AVERage OFF|<int>
CONFigure:CONDition:AVERage?
This command defines whether or not to activate the averaging function. The
choices are OFF and <int> for the number of averaging times. The allowable range
of <int> is from 2 to 99. At preset, this is set to OFF.
BER BS Level
CONFigure:CONDition:SENSitivity:LEVel <real> [DBM]
CONFigure:CONDition:SENSitivity:LEVel?
This command specifies an amplitude of the Test Set for error ratio tests, for each
radio standard set by the Radio Format command. The allowable range is from
110.0 to 50.0 (dBm) in 0.5 steps. At preset, this is set to 102.0 (dBm) for
GSM900 and E-GSM, 100.0 (dBm) for DCS1800, or 102.0 (dBm) for
PCS1900.
BER Frames
CONFigure:CONDition:SENSitivity:FRAMe <int>
CONFigure:CONDition:SENSitivity:FRAMe?
This command specifies the number of the Test Set for error ratio tests. The
allowable range is from 1 to 13000 (frames) in 1 steps. At preset, this is set to 10
(frames).
BER Limit
CONFigure:CONDition:LIMit:BER <real h> [PCT]
CONFigure:CONDition:LIMit:BER?
This command specifies the high limit for BER (bit error ratio) tests. The
allowable range is from 0.00 to 99.99 (%) in 0.01 steps. At preset, this is set to 2.44
(%).