Specifications
Table Of Contents
- E6392-90042.pdf
- E6392-90042_manual
- Front Matter
- Contents
- 1 Preparing for Use
- 2 Programming Command Guidelines
- 3 Programming Command Reference
- IEEE Common Command
- CONFigure Subsystem
- DISPlay Subsystem
- HCOPy Subsystem
- RFGenerator Subsystem
- SMONitor Subsystem
- SYSTem Subsystem
- TESTs Subsystem for AUTOMATIC TEST
- All Tests Results
- All Tests Results
- Command Reference for Burst Timing/Power Ramp Measurement Screen
- Command Reference for DC Current Measurement Screen
- Command Reference for MS Information
- Command Reference for Peak TX Power Measurement Screen
- Command Reference for Phase/Frequency Error Measurement Screen
- Command Reference for Sensitivity/RX Quality/RX Level Measurement Screen
- Command Reference for Stand-by/Measuring Screens
- TESTs Subsystem for MANUAL TEST
- Command Reference for DC Current Measurement Screen
- Command Reference for Measuring Screens
- Command Reference for MS Information
- Command Reference for Peak TX Power /Burst Timing/Power Ramp Measurement Screen
- Command Reference for Phase/Frequency Error Measurement Screen
- Command Reference for Sensitivity/RX Quality/RX Level Measurement Screen
- Command Reference for Spectrum Monitor Measurement Screen
- Command Reference for Stand-by Screens
- TRIGger Subsystem
- 4 Programming Command Cross Reference
- 5 Example Programs
- A Syntax Diagram
- B Command Difference between E6392A & E6392B
- Index

34 Chapter 3
Programming Command Reference
IEEE Common Commands
IEEE Common Commands
The following IEEE common commands are used to set and monitor the status
registers, and to reset the Test Set.
IEEE Common Commands Reference
Clear Status
*CLS
This command initializes the status byte register (STBR) and the standard event
status register (SESR).
Event Status Register Query
*ESR?
This query command returns a value of the standard event status register (SESR).
The range of a value is from 0 to 255.
Identification Query
*IDN?
This query command returns a string of “AGILENT TECHNOLOGIES,
E6392B, <serial number>, <firmware revision number>”.
Instrument Option Query
*OPT?
This query command returns a value of the option numbers installed in the Test Set
as follows:
000 No option
002 Asynchronous Test Mode for MANUAL TEST
040 GPRS Test Mode for MANUAL TEST
Operation Complete
*OPC
*OPC?
This command enables or queries the OPC bit of the standard event status register
(SESR). The query command returns 1.
Reset
*RST
This command executes to run the selftest routine and resets the Test Set to the
initial state.