Specifications
Table Of Contents
- E6392-90042.pdf
- E6392-90042_manual
- Front Matter
- Contents
- 1 Preparing for Use
- 2 Programming Command Guidelines
- 3 Programming Command Reference
- IEEE Common Command
- CONFigure Subsystem
- DISPlay Subsystem
- HCOPy Subsystem
- RFGenerator Subsystem
- SMONitor Subsystem
- SYSTem Subsystem
- TESTs Subsystem for AUTOMATIC TEST
- All Tests Results
- All Tests Results
- Command Reference for Burst Timing/Power Ramp Measurement Screen
- Command Reference for DC Current Measurement Screen
- Command Reference for MS Information
- Command Reference for Peak TX Power Measurement Screen
- Command Reference for Phase/Frequency Error Measurement Screen
- Command Reference for Sensitivity/RX Quality/RX Level Measurement Screen
- Command Reference for Stand-by/Measuring Screens
- TESTs Subsystem for MANUAL TEST
- Command Reference for DC Current Measurement Screen
- Command Reference for Measuring Screens
- Command Reference for MS Information
- Command Reference for Peak TX Power /Burst Timing/Power Ramp Measurement Screen
- Command Reference for Phase/Frequency Error Measurement Screen
- Command Reference for Sensitivity/RX Quality/RX Level Measurement Screen
- Command Reference for Spectrum Monitor Measurement Screen
- Command Reference for Stand-by Screens
- TRIGger Subsystem
- 4 Programming Command Cross Reference
- 5 Example Programs
- A Syntax Diagram
- B Command Difference between E6392A & E6392B
- Index

Chapter 3 107
Programming Command Reference
TESTs Subsystem for MANUAL TEST
3. Programming Command
Reference
DC Power Voltage
TESTs:MANual:SMONitor:DCPower:VOLT <real> [V]
TESTs:MANual:SMONitor:DCPower:VOLT?
This command specifies a voltage value of DC Power when the DC Power supply
mode in the Stand-by screen is set to ON. The allowable range is from 3.0 to 11.0
(V) in 0.1 steps. At preset, this is set to 3.0 (V).
Marker Measurement
TESTs:MANual:SMONitor:RESults:MARKer?
This query command returns a set of values, <string 1>, <string 2>,
<string 3>, <string 4>, and <string 5>, for the marker position.
• <string 1> (1 digit) is 0 when all values are within the normal range, or 1 if any
measurement is overranged, underranged or not made.
• The definitions and effective ranges of <string 2> (10 digits), <string 3> (5
digits), <string 4> (7 digits), and <string 5> (5 digits) are shown in the table
below. If any measurement is overranged, underranged or not made, the values
9999999999 for <string 2>, −99.9 for <string 3> and <string 5>, and +999999
for <string 4>, are returned.
Marker Position Control
TESTs:MANual:SMONitor:RESults:MARKer:POINt UP|DOWN
TESTs:MANual:SMONitor:RESults:MARKer:POINt <int> [HZ]
TESTs:MANual:SMONitor:RESults:MARKer:POINt?
This command enables you to move the marker position UP or DOWN in steps, or
by <int> in Hz ranging from 0 to 400000. At preset, this is set to 1000 (Hz).
Measurement Result
TESTs:MANual:SMONitor:RESults?
This query command returns a set of test result values, <string 1>,
<string 2>, <string 3>, <string 4> and <string 5>.
Value Definition Range
<string 2> Absolute RF frequency [Hz] 0880200000 to 0915200000,
1710200000 to 1785200000,
1850200000 to 1909800000
<string 3> Absolute RF level [dBm] −99.9 to +99.9
<string 4> Relative RF frequency [Hz] +000000 to +400000
<string 5> Relative RF level [dB] −99.9 to +0.00