Specifications
Table Of Contents
- E6392-90042.pdf
- E6392-90042_manual
- Front Matter
- Contents
- 1 Preparing for Use
- 2 Programming Command Guidelines
- 3 Programming Command Reference
- IEEE Common Command
- CONFigure Subsystem
- DISPlay Subsystem
- HCOPy Subsystem
- RFGenerator Subsystem
- SMONitor Subsystem
- SYSTem Subsystem
- TESTs Subsystem for AUTOMATIC TEST
- All Tests Results
- All Tests Results
- Command Reference for Burst Timing/Power Ramp Measurement Screen
- Command Reference for DC Current Measurement Screen
- Command Reference for MS Information
- Command Reference for Peak TX Power Measurement Screen
- Command Reference for Phase/Frequency Error Measurement Screen
- Command Reference for Sensitivity/RX Quality/RX Level Measurement Screen
- Command Reference for Stand-by/Measuring Screens
- TESTs Subsystem for MANUAL TEST
- Command Reference for DC Current Measurement Screen
- Command Reference for Measuring Screens
- Command Reference for MS Information
- Command Reference for Peak TX Power /Burst Timing/Power Ramp Measurement Screen
- Command Reference for Phase/Frequency Error Measurement Screen
- Command Reference for Sensitivity/RX Quality/RX Level Measurement Screen
- Command Reference for Spectrum Monitor Measurement Screen
- Command Reference for Stand-by Screens
- TRIGger Subsystem
- 4 Programming Command Cross Reference
- 5 Example Programs
- A Syntax Diagram
- B Command Difference between E6392A & E6392B
- Index

Chapter 3 103
Programming Command Reference
TESTs Subsystem for MANUAL TEST
3. Programming Command
Reference
RX Level Measurement
TESTs:MANual:RX:RESults:LEVel?
This query command returns a set of values, <string 1>, <string 2> and
<string 3> for RX Level tests.
• <string 1> (1 digit) is 0 when all values are within the normal range, or 1 if any
measurement is overranged, underranged or not made.
• <string 2> (1 digit) is the binary data, 0 or 1, for the pass (0) or fail (1) result.
• <string 3> (2 digits) is the measured RX Level factor ranging from 00 (<−110
dBm) to 63 (>−48 dBm). If its measurement is overranged, underranged or not
made, the value 63 is returned.
NOTE If the mode is set to GPRS on the Test Set with Option 040, this
query command is invalid.
RX Quality Measurement
TESTs:MANual:RX:RESults:QUALity?
This query command returns a set of values, <string 1>, <string 2> and
<string 3>, for RX Quality tests.
• <string 1> (1 digit) is 0 when all values are within the normal range, or 1 if any
measurement is overranged, underranged or not made.
• <string 2> (1 digit) is the binary data, 0 or 1, for the pass (0) or fail (1) result.
• <string 3> (1 digit) is the measured RX Quality factor ranging from 0 (<0.2%)
to 7 (>12.8%). If its measurement is overranged, underranged or not made, the
value 7 is returned.
NOTE If the mode is set to GPRS on the Test Set with Option 040, this
query command is invalid.
RFORmat Allowable Range Preset Value
GSM900 2 (+39 dBm) to 19 (+5 dBm) 5 (+39 dBm)
E-GSM 2 (+39 dBm) to 19 (+5 dBm) 5 (+39 dBm)
DCS1800 0 (+30 dBm) to 15 (+0 dBm), or
29 (+36 dBm) to 31 (+32 dBm)
0 (+39 dBm)
PCS1900 0 (+30 dBm) to 15 (+0 dBm), or
30 (+33 dBm) to 31 (+32 dBm)
0 (+39 dBm)