102 Chapter 3
Programming Command Reference
TESTs Subsystem for MANUAL TEST
Command Reference for Sensitivity/RX Quality/RX Level Measurement
Screen
The SENSitivity commands are used to set the controls and parameters associated
with the Sensitivity/RX Quality/RX Level measurements and the pass/fail tests.
Send the "DISPlay:MANual:SENSitivity" command to display the Sensitivity/RX
Quality/RX Level Measurement screen for MANUAL TEST before sending the
following commands.
NOTE If the mode is set to ASYNC on the Test Set with Option 002, the
commands related to Sensitivity/RX Quality/RX Level
measurements and pass/fail tests are invalid.
NOTE If the mode is set to GPRS on the Test Set with Option 040, the
commands related to RX Quality and RX Level measurements and
their pass/fail tests are invalid.
DC Power Voltage
TESTs:MANual:SENSitivity:DCPower:VOLT <real> [V]
TESTs:MANual:SENSitivity:DCPower:VOLT?
This command specifies a voltage value of DC Power when the DC Power supply
mode in the Stand-by screen is set to ON. The allowable range is from 3.0 to 11.0
(V) in 0.1 steps. At preset, this is set to 3.0 (V).
Power Control Level
TESTs:MANual:SENSitivity:PLEVel <int>
TESTs:MANual:SENSitivity:PLEVel?
This command specifies a value of Power Control Level for each radio standard set
by the Radio Format command. The allowable range and preset values are as
follows:
DCS1800 GSM900 1710200000 to 1784800000;
890200000 to 914800000
1710200000
E-GSM 1710200000 to 1784800000;
880200000 to 914800000
OFF 1710200000 to 1784800000
PCS1900 OFF 1850200000 to 1909800000 1850200000
RFORmat MBANd Allowable Range [Hz] Preset Value [Hz]