User manual

Register Definitions B
E6198B Switch/Load Unit User Manual B-15
Agilent E6176A 16-Channel High-Current Load Card
The Agilent E6176A Load Card is designed for high-current
loads mounted outside the load card. The card provides current
sense, pull-up/down, and flyback protection. The card also
provides Card Type, Card Configuration, and Status readback of
the built in relay timers. See “Using the Agilent E6176A
16-Channel Load Card" for more information.
Register definitions for the card follow:
Card Type (R) Base + 0
h
This register reads back the Card Type (02
h
) of the card.
Read Only
Power On/Reset State = 02
h
Card Configuration (R) Base + 1
h
This register reads back the Card Configuration. The Card
Configuration is determined by the user to distinguish different
load configurations of the same load card type.
Read Only
Undefined bits readback as all 1s
Status (R) Base + 2
h
This register reads back the card's status. Currently the status is
defined as the state of the relay timer. The relay timer has two
timers wire-OR'd together. One timer is designed for the slower
armature relays (>16 ms) and the second designed for the faster
reed relays (>500 μs). The timers restart whenever the registers
Bit 7-0
Purpose
Card Type
Setting 02
h
Bit 7-0
Purpose
Card Configuration
Setting state