Using Load Cards and Loads 5
E6198B Switch/Load Unit User Manual 5-35
External Load Mounting Options
Loads typically consist of “real” devices (injectors, actuators,
etc.) that cannot be mounted on the load card. If cable
resistance is a significant factor (such as for low form-factor or
low-value resistive loads), one solution is to mount the loads on
a metal plate that fits into the SLU slots adjacent to the load
card (Figure 5-25). Follow these guidelines if you choose to
mount the loads externally:
Mount the loads as close to the load card as possible.
The dimensions of the load plate for the 16-channel card are:
Height- 23.32 cm (9.18 in.)
Length (Max)- 34.00 cm (13.386 in.)
Thickness- 1.59 mm (1/16 in.).
Arrange the loads so that the lowest impedances have the
shortest cable runs.
Route cable runs to the externally mounted loads so that you
can remove other cards without having to disconnect any
wiring.
Clearly identify all cables to externally mounted loads,
including those on a load plate. Use labels, numbers, color
coding, or some combination of these methods. Cabling to
polarity sensitive devices should be labeled appropriately.