Specifications
Diagnostic Testing Details 4
TS-8900 Functional Test System Diagnostics User Guide 4-65
CEDGN_M9186A_VI.tpa (VI Source CEDGN Test Flow)
The testsplan supports testing of two M9186A VI Source
modules.
It is used to verify the connectivity of VI Source module Safety
Interlock on the Express Connect TC connector. Test will abort
with EXCEPTION ENCOUNTERED error if safety interlock is
not installed.
User is required to set the variables to reflect the system under
test.
1 = present, 0 = not present
bVI1 = 1st VI Source M9186A
bVI2 = 2nd VI Source M9186A
Device that user sets as not present will not be tested. The test
starts by performing:
Not Acceptable Combination Checking
• Checks user settings for any not acceptable combination. Stops
test and provide error message if not acceptable selection is found.
M9186A VI Source Module VI1 Tests
• Display instruction to install CEDGN Board #3 on appropriate TC.
• Setup DMM to perform voltage measurement covering:
VI1 Low Voltage 0ohm Path
VI1 Low Voltage 10ohm Path
VI1 Low Voltage 100ohm Path
VI1 Low Voltage 1kohm Path
VI1 Low Voltage 10kohm Path
VI1 High Voltage 0ohm Path
VI1 High Voltage 100ohm Path
• Setup DMM to perform current measurement covering:
VI1 Low Current
VI1 High Current