Specifications

Diagnostic Testing Details 4
TS-8900 Functional Test System Diagnostics User Guide 4-61
CH5 with/without sense
CH6 with/without sense
CH7 with/without sense
CH8 with/without sense
Display instruction to toggle S1 and S2 on Board #2 to UP position.
Setup PXI DAC8_1 to output 16V on CH8 when External Trigger is
detected.
Use SLU internal DAC1 to generate a simulated trigger to PXI
DAC8_1.
Measure for 16V at PXI DAC8_1 CH8 output for trigger response.
M9185A 8 Channel PXI DAC2 (CH1-CH8) Tests
Display instruction to install CEDGN Board #1 on TC1. Install
CEDGN Board #2 on appropriate TC and toggle S1 and S2 on Board
#2 to DOWN position. Install Cable E6170-61607 on J1 of Board #1
to J1 of Board #2.
Setup DMM to perform voltage measurement.
Measure DAC8_2 output for (-16V, +16V and 0V) on channel:
CH1 with/without sense
CH2 with/without sense
CH3 with/without sense
CH4 with/without sense
CH5 with/without sense
CH6 with/without sense
CH7 with/without sense
CH8 with/without sense
Display instruction to toggle S1 and S2 on Board #2 to UP position.
Setup PXI DAC8_2 to output 16V on CH8 when External Trigger is
detected.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com