Specifications
Diagnostic Testing Details 4
TS-8900 Functional Test System Diagnostics User Guide 4-59
M9185A 16 Channel PXI DAC1 (CH1-CH8) on TC1 Tests
• Display instruction to install CEDGN Board #2 on TC1 and toggle
S1 and S2 on Board #2 to DOWN position.
• Setup DMM to perform voltage measurement.
• Measure DAC16_1 output for (-16V, +16V and 0V) on channel:
CH1 with/without sense
CH2 with/without sense
CH3 with/without sense
CH4 with/without sense
CH5 with/without sense
CH6 with/without sense
CH7 with/without sense
CH8 with/without sense
• Display instruction to toggle S1 and S2 on Board #2 to UP position.
• Setup PXI DAC16_1 to output 16V on CH8 when External Trigger is
detected.
• Use SLU internal DAC1 to generate a simulated trigger to PXI
DAC16_1.
• Measure for 16V at PXI DAC16_1 CH8 output for trigger response.
M9185A 16 Channel PXI DAC1 (CH9-CH16) on TC3 Tests
• Display instruction to install CEDGN Board #1 on TC1. Install
CEDGN Board #2 on TC3 and toggle S1 and S2 on Board #2 to
DOWN position. Install Cable E6170-61607 on J1 of Board #1 to J1
of Board #2.
• Setup DMM to perform voltage measurement.
• Measure DAC16_1 output for (-16V, +16V and 0V) on channel:
CH9 with/without sense
CH10 with/without sense
CH11 with/without sense
CH12 with/without sense
CH13 with/without sense
CH14 with/without sense
CH15 with/without sense
CH16 with/without sense
M9185A 16 Channel PXI DAC2 (CH1-CH8) on TC5 Tests
• Display instruction to install CEDGN Board #1 on TC1. Install
CEDGN Board #2 on TC5 and toggle S1 and S2 on Board #2 to
DOWN position. Install Cable E6170-61607 on J1 of Board #1 to J1
of Board #2.
• Setup DMM to perform voltage measurement.
• Measure DAC16_2 output for (-16V, +16V and 0V) on channel:
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com