Specifications

4-58 TS-8900 Functional Test System Diagnostics User Guide
4 Diagnostic Testing Details
CEDGN_M9185A_DAC.tpa (8/16 Channel PXI DAC CEDGN Test Flow)
The testsplan is for testing of M9185A 8/16 Channel DAC. It
supports two 16 Channel DAC or three 8 Channel DAC or one 16
Channel and one 8 Channel DAC.
User is required to set the variables to reflect the system under
test.
1 = present, 0 = not present
bDAC16_1 = 1st 16 Channel M9185A
bDAC16_2 = 2nd 16 Channel M9185A
bDAC8_1 = 1st 8 Channel M9185A
bDAC8_2 = 2nd 8 Channel M9185A
bDAC8_2 = 3rd 8 Channel M9185A
Device that user sets as not present will not be tested. The test
starts by performing:
Not Acceptable Combination Checking
Checks user settings for any not acceptable combination. Stops
test and provide error message if not acceptable selection is found.