Specifications

4-52 TS-8900 Functional Test System Diagnostics User Guide
4 Diagnostic Testing Details
CEDGN_E6198B_Utility.tpa (Switch Load Unit Utility CEDGN Test Flow)
The testsplan performs testing on the SLU Utility Connector at
the Express Connect front panel.
SLU Utility Setup Information
Display instruction to install CEDGN Board #1 on TC1. Install Cable
E6170-61607 on SLU Utility Connector to J2 of Board #1.
SLU Utility System_Gnd Tests
Setup DMM to measure voltage.
Measure for 12V using fixed +12V reference at pin49 of SLU Utility
Connector and changing System_Gnd. Covers System_Gnd at pin1,
pin2, pin11, pin12, pin55, pin56, pin57, pin58, pin62 and pin64.
SLU Utility +12Vdc_Supply Tests
Setup DMM to measure voltage.
Measure for 12V using fixed System_Gnd at pin1 of SLU Utility
Connector and changing +12Vdc_Supply. Covers +12Vdc_Supply at
pin49, pin50, pin51 and pin52.
SLU Utility -12Vdc_Supply Tests
Setup DMM to measure voltage.
Measure for -12V between System_Gnd at pin1 and -12Vdc_Supply
at pin54 of SLU Utility Connector.
SLU Utility DAC1 Tests
Setup DMM to measure voltage.
Measure for DAC1 output voltage at -16V, +16V and 0V.
SLU Utility DAC2 Tests
Setup DMM to measure voltage.
Measure for DAC2 output voltage at -16V, +16V and 0V.