Specifications
Diagnostic Testing Details 4
TS-8900 Functional Test System Diagnostics User Guide 4-33
U7177A 24 Channel Load Card LC24CH_4 Tests
• Display instruction to Install load module E2240-67012 on
LC24CH_4 Load Card.
• Display instruction to install CEDGN Board #1 on TC1 and CEDGN
Board #4 on appropriate TC. Install Cable E6170-61607 from SLU
Utility Connector to J2 of Board #4. Install Cable E6170-61607 from
J1 of Board #1 to J1 of Board #4.
• Setup DMM to perform 2W resistance measurement.
• Measure LC24CH_4 load card loopback path resistance from
CH1-Com1 until CH12-Com12 via Row1 and Row8
CH13-Com13 until CH24-Com24 via Row2 and Row7
• Measure LC24CH_4 load card sense resistor path:
CH1 Sense until CH24 Sense via Row25 and Row26.
U7177A 24 Channel Load Card LC24CH_5 Tests
• Display instruction to Install load module E2240-67012 on
LC24CH_5 Load Card.
• Display instruction to install CEDGN Board #1 on TC1 and CEDGN
Board #4 on appropriate TC. Install Cable E6170-61607 from SLU
Utility Connector to J2 of Board #4. Install Cable E6170-61607 from
J1 of Board #1 to J1 of Board #4.
• Setup DMM to perform 2W resistance measurement.
• Measure LC24CH_5 load card loopback path resistance from
CH1-Com1 until CH12-Com12 via Row1 and Row8
CH13-Com13 until CH24-Com24 via Row2 and Row7
• Measure LC24CH_5 load card sense resistor path:
CH1 Sense until CH24 Sense via Row25 and Row26.