Specifications
4-28 TS-8900 Functional Test System Diagnostics User Guide
4 Diagnostic Testing Details
• Measure LC8CH_4 load card sense resistor path:
CH1 & CH2 Current Sense, CH3 & CH4 Current Sense, CH5 & CH6
Current Sense, and CH7 & CH8 Current Sense via Row25 and
Row26.
E6175A 8 Channel Load Card LC8CH_5 Tests
• Display instruction to Install load module E2240-67012 on LC8CH_5
Load Card.
• Display instruction to install CEDGN Board #1 on TC1 and CEDGN
Board #4 on appropriate TC. Install Cable E6170-61607 from SLU
Utility Connector to J2 of Board #4. Install Cable E6170-61607 from
J1 of Board #1 to J1 of Board #4.
• Setup DMM to perform 2W resistance measurement.
• Measure LC8CH_5 load card loopback path resistance from
CH1-CH5, CH2-CH6, CH3-CH7 and CH4-CH8 via Row1 and Row2.
• Measure LC8CH_5 load card sense resistor path:
CH1 & CH2 Current Sense, CH3 & CH4 Current Sense, CH5 & CH6
Current Sense, and CH7 & CH8 Current Sense via Row25 and
Row26.
E6175A 8 Channel Load Card LC8CH_6 Tests
• Display instruction to Install load module E2240-67012 on LC8CH_6
Load Card.
• Display instruction to install CEDGN Board #1 on TC1 and CEDGN
Board #4 on appropriate TC. Install Cable E6170-61607 from SLU
Utility Connector to J2 of Board #4. Install Cable E6170-61607 from
J1 of Board #1 to J1 of Board #4.
• Setup DMM to perform 2W resistance measurement.
• Measure LC8CH_6 load card loopback path resistance from
CH1-CH5, CH2-CH6, CH3-CH7 and CH4-CH8 via Row1 and Row2.
• Measure LC8CH_6 load card sense resistor path:
CH1 & CH2 Current Sense, CH3 & CH4 Current Sense, CH5 & CH6
Current Sense, and CH7 & CH8 Current Sense via Row25 and
Row26.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com