Specifications

4-24 TS-8900 Functional Test System Diagnostics User Guide
4 Diagnostic Testing Details
Matrix4 E8783A Row33-Row64 Tests
Setup DMM for 4W resistance measurement.
Perform Matrix4 loopback resistance measurements between
Row33-Row34, Row35-Row36, until Row63-Row64.
Matrix5 Row1-Row32
Display instruction to Install CEDGN Board #5 on TC8. S1 and S2
on CEDGN Board #5 are all toggled to ON (CLOSE).
Matrix5 E8783A Row1-Row32 Tests
Setup DMM for 4W resistance measurement.
Perform Matrix5 loopback resistance measurements between
Row1-Row2, Row3-Row4, until Row31-Row32.
Matrix 1 Aux 1-8 Testing
Display instruction to Install CEDGN Board #2 on TC1. S1 and S2
on CEDGN Board #2 are toggle to DOWN position.
Setup DMM for 4W resistance measurement.
Perform Matrix1 loopback resistance measurements between
Aux1-Aux2, Aux3-Aux4, Aux5-Aux6 and Aux7-Aux8.
Matrix 1 Aux 33-40 Testing
Display instruction to Install CEDGN Board #2 on TC3. S1 and S2
on CEDGN Board #2 are toggle to DOWN position.
Setup DMM for 4W resistance measurement.
Perform Matrix1 loopback resistance measurements between
Aux33-Aux34, Aux35-Aux36, Aux37-Aux38 and Aux39-Aux40.
Matrix 2 Aux 1-8 Testing
Display instruction to Install CEDGN Board #2 on TC5. S1 and S2
on CEDGN Board #2 are toggle to DOWN position.
Setup DMM for 4W resistance measurement.
Perform Matrix2 loopback resistance measurements between
Aux1-Aux2, Aux3-Aux4, Aux5-Aux6 and Aux7-Aux8.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com