Specifications

3-12 TS-8900 Functional Test System Diagnostics User Guide
3 Running Diagnostics
Resolving Test Failures
Test failures can be caused by improper switch settings on the
Diagnostic Test Fixture, improper system equipment
specification, or the actual test system instrument failure.
Before troubleshooting the system equipment, make sure the
test fixture switches are properly set and the test system
instrumentation is properly specified. If the problem appears to
be system equipment, swap the instrument and/or the cable
between the instrument and the Test System Interface.
Chapter 4 contains detailed descriptions of each test that may
help in isolating equipment problems to a particular
component.