Specifications

3-6 TS-8900 Functional Test System Diagnostics User Guide
3 Running Diagnostics
C:\Program Files\Agilent\TS-5000 System Software\Service\
U8970A\Ust
4 Load the topology file (*. ust) of the hardware to be tested
and edit the Modules folder content:
Disable - Check to disable. Uncheck to enable. Depends on
availability of the module in the system.
Set "Slot #" - Depends on the slot number the card is
installed on the Switch Load Unit. (Applicable to SLU
cards only)
Set "Instrument Descriptor/Resource" - LXI, PXI, GPIB
module address. Obtained from Agilent IO Libraries Suite.
(VISA Address)
Set "PCI Slot Number" and "Serial Number" - Depends on
the installed Softing CAN.
Once complete editing, save and close the topology file.
5 Load the testplan together with its corresponding topology
file. Before running the testplan, identify the
instrumentation configuration you want to test in the first
lines of the testplan as shown on Figure 3-2. Place a 1 next to
instrument configuration you want to test, a 0 next to those
instrument configurations you do not want to test (or that
are not in your system).
6 Click to run the testplan. The testplan will run tests on
each system instrument configuration specified in step 5 and
display the results of the testing.
7 Follow the instructions appearing on the display throughout
the test.
8 Repeat step 4 until all instrumentation available in your
system are tested.