4-78 TS-8900 Functional Test System Diagnostics User Guide
4 Diagnostic Testing Details
CEDGN_SoftingCAN.tpa (Softing CAN CEDGN Test Flow)
The testsplan supports testing of two Softing CAN cards. Each
card has two CAN ports. User is required to set the variables to
reflect the system under test.
1 = present, 0 = not present
bCAN1_1_AND_CAN1_2_TC1 = 1st Softing CAN card with CAN
ports on TC1
bCAN2_1_TC3 = 2nd Softing CAN card with CAN port 1 on TC3
bCAN2_2_TC5 = 2nd Softing CAN card with CAN port 2 on TC5
Device that user sets as not present will not be tested. The test
starts by performing:
CAN1_1 and CAN1_2 communication test
Display instruction to Install CEDGN Board #1 on TC1.
Use CAN1_1 to transmit to CAN1_2. Verify transmitted data
matched received data.
Use CAN1_2 to transmit to CAN1_1. Verify transmitted data
matched received data.
CAN1_1 and CAN2_1 communication test
Display instruction to Install CEDGN Board #1 on TC1 and CEDGN
Board #2 on TC3. Install Cable E6170-61607 on J1 of Board #1 to
J1 of Board #2.
Use CAN1_1 to transmit to CAN2_1. Verify transmitted data
matched received data.
Use CAN2_1 to transmit to CAN1_1. Verify transmitted data
matched received data.
CAN1_1 and CAN2_2 communication test
Display instruction to Install CEDGN Board #1 on TC1 and CEDGN
Board #2 on TC5. Install Cable E6170-61607 on J1 of Board #1 to
J1 of Board #2.
Use CAN1_1 to transmit to CAN2_2. Verify transmitted data
matched received data.
Use CAN2_2 to transmit to CAN1_1. Verify transmitted data
matched received data.