Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment • FAST SHIPPING AND DELIVERY • TENS OF THOUSANDS OF IN-STOCK ITEMS • EQUIPMENT DEMOS • HUNDREDS OF MANUFACTURERS SUPPORTED • LEASING/MONTHLY RENTALS • ITAR CERTIFIED SECURE ASSET SOLUTIONS SERVICE CENTER REPAIRS Experienced engineers and technicians on staff at our full-service, in-house repair center WE BUY USED EQUIPMENT Sell your excess, underutilized, and idle used equipment We also offer credit for buy-back
Agilent TS-8900 Functional Test System Diagnostics User Guide Agilent Technologies Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
Notices © Agilent Technologies, Inc. 2011, 2012 Manual Part Number No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws. U8970-90003 Edition Second Edition, November 2012 Printed in Malaysia Agilent Technologies Microwave Products (Malaysia) Sdn. Bhd.
Table Of Content 1 Legal Information Legal Information 1-2 Warranty 1-2 Technology Licenses 1-2 Restricted Rights Legend 1-2 Service And Support 1-3 Agilent On The Web 1-3 Agilent By Phone 1-3 2 Safety and Regulatory Information Safety Information 2-2 Safety Summary 2-2 Safety Notice 2-2 General 2-2 Environmental Conditions 2-3 Before Applying Power 2-3 Ground The System 2-4 Fuses 2-4 Operator Safety Information 2-4 Safety Symbols and Regulatory Markings Electrostatic Discharge (ESD) Precautions 2-5 2-7
U8970-66606 PCA-DIAGNOSTIC NUMBER 4 (Board #4) 4-17 U8970-66607 PCA-DIAGNOSTIC NUMBER 5 (Board #5) 4-17 E6170-61607 CABLE 2X 2X32 IDC PIN 1 TO PIN 1 4-18 E6170-61618 Diagnostic Cable 4-18 E2240-67012 CEDGN LOOPBACK CONNECTOR 4-19 E6170-66520 820OHM LOADED MEZZANINE CARD 4-19 E6170-61619 CABLE-HEAVY DUTY LOAD JUMPER 4-20 CEDGN Testplan Description and Flow 4-21 CEDGN_E878XA_PinMatrix.tpa (Pin Matrix CEDGN Test Flow) 4-22 CEDGN_E6178B_LoadCard.
List of Figure 1 Legal Information 2 Safety and Regulatory Information 3 Running Diagnostics Figure 3-1. Specify the Instruments to be Tested in DGN Testplan 3-4 Figure 3-2. Specify the Instruments to be Tested in CEDGN Testplan 3-7 Figure 3-3. Installing CEDGN Board #1 (U8970-66603) to TC1 and CEDGN Board #3 (U8970-66605) to TC4 with both boards linked by cable (E6170-61607). 3-9 Figure 3-4.
Figure 4-23. E6198B Switch Load Unit Utility Test 4-54 Figure 4-24. L4451A LXI DAC Test 4-57 Figure 4-25. M9185A PXI DAC Output Test 4-63 Figure 4-26. M9185A PXI DAC Trigger Test 4-64 Figure 4-27. M9187A DIO Test 4-69 Figure 4-28. M9216A High Voltage Data Acquisition Test A 4-75 Diagnostic Test Fixture Details Figure A-1. U8970-66603 PCA-Diagnostic Number 1 (Board #1) A-2 Figure A-2. U8970-66604 PCA-Diagnostic Number 2 (Board #2) A-3 Figure A-3.
List of Table 1 Legal Information Table 1-1. Agilent Call Centers and Regional Headquarters 2 1-3 Safety and Regulatory Information Table 2-1. Environment Requirements 2-3 Table 2-2. Safety Symbols and Regulatory Markings 2-5 Table 2-3. Suggested Anti-Static Solutions for Site Planning 3 Running Diagnostics Table 3-1. CEDGN Testplan Description and Flow 4 3-5 Diagnostic Testing Details Table 4-1. CEDGN Testplan Description and Flow A 2-7 4-21 Diagnostic Test Fixture Details Table A-1.
THIS PAGE IS INTENTIONALLY LEFT BLANK. viii TS-8900 Functional Test System Wiring Guide and Hardware Reference Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
Agilent TS-8900 Functional Test Sytem Diagnostics User Guide 1 Legal Information Warranty 1-2 Technology Licenses 1-2 Restricted Rights Legend 1-2 Service And Support 1-3 Agilent On The Web 1-3 Agilent By Phone 1-3 Agilent Technologies Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
1 Legal Information Legal Information Warranty The material contained in this document is provided “as is, ” and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose.
Legal Information 1 Service And Support Any adjustment, maintenance, or repair of this product must be performed by qualified personnel. Contact your customer engineer through your local Agilent Technologies Service Center. Agilent On The Web You can find information about technical and professional services, product support, and equipment repair and service on the Web: http://www. agilent. com/ Double-click the link to Test & Measurement. Select your country from the drop-down menus.
1 Legal Information Manufacturing Address Agilent Technologies Microwave Products (Malaysia) Sdn. Bhd. Bayan Lepas Free Industrial Zone, 11900 Penang, Malaysia. 1-4 TS-8900 Functional Test System Diagnostics User Guide Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
Agilent TS-8900 Functional Test System Diagnostics User Guide 2 Safety and Regulatory Information Safety Information 2-2 Safety Summary 2-2 Safety Notice 2-2 General 2-2 Environmental Conditions 2-3 Before Applying Power 2-3 Ground The System 2-4 Fuses 2-4 Operator Safety Information 2-4 Safety Symbols and Regulatory Markings 2-5 Electrostatic Discharge (ESD) Precautions 2-7 End of Life: Waste Electrical and Electronic Equipment (WEEE) Directive 2002/96/EC 2-8 Agilent Technologies Artisan Technology Grou
2 Safety and Regulatory Information Safety Information Safety Summary The following general safety precautions must be observed during all phases of operation of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Agilent Technologies, Inc. assumes no liability for the customer's failure to comply with these requirements.
2 Safety and Regulatory Information WA RNING DO NOT REMOVE RACK PANELS OR INSTRUMENT COVERS. Operating personnel must not remove any rack panels or instrument covers. Component replacement and internal adjustments must be made only by qualified service personnel. Products that appear damaged or defective should be made inoperative and secured against unintended operation until they can be repaired by a qualified service personnel.
2 Safety and Regulatory Information Ground The System To minimize shock hazard, the instrument chassis and cover must be connected to an electrical protective earth ground. The instrument must be connected to the ac power mains through a grounded power cable, with the ground wire firmly connected to an electrical ground (safety ground) at the power outlet.
2 Safety and Regulatory Information cables to the front panel connectors of the module through non-conductive, flexible conduit such as that used in electrical power distribution. Safety Symbols and Regulatory Markings Symbols and markings on the system, in manuals and on instruments alert you to potential risks, provide information about conditions, and comply with international regulations. Table 2-2 defines the symbols and markings you may find in a manual or on an instrument.
2 Safety and Regulatory Information Safety symbols Terminal for line conductor on permanently installed equipment. Standby (supply); units with this symbol are not completely disconnected from ac mains when this switch is off. To completely disconnect the unit from ac mains, either disconnect the power cord, or have a qualified electrician install an external switch. Regulatory markings This text indicates that the ISM device complies with Canadian ICES-001.
Safety and Regulatory Information 2 Electrostatic Discharge (ESD) Precautions Static electricity is destructive to your production process and the TS-8900. Careless handling and poor site planning can cause system reliability problems and reduce your product yield. The system may not be as easily damaged as the modules you will be testing, but good anti-static planning will help ensure high reliability. The ESD symbol below indicates areas where ESD caution must be exercised.
2 Safety and Regulatory Information End of Life: Waste Electrical and Electronic Equipment (WEEE) Directive 2002/96/EC This product complies with the WEEE Directive (2002/96/EC) marking requirement. The affixed product label (see below) indicates that you must not discard this electrical/electronic product in domestic household waste. Product Category: With reference to the equipment types in the WEEE directive Annex 1, this product is classified as a “Monitoring and Control Instrumentation” product.
Agilent TS-8900 Functional Test System Diagnostics User Guide 3 Running Diagnostics Overview of Diagnostics Testplans 3-2 Configuring and Running the DGN Testplan 3-3 Configuring and Running the CEDGN Testplan 3-5 Installing the Diagnostic Test Fixture 3-8 Resolving Test Failures 3-12 This chapter describes the overview of diagnostic testplans, DGN testplan, CEDGN testplan and how to use the diagnostics test fixtures to verify the operation of system equipment.
3 Running Diagnostics Overview of Diagnostics Testplans The verification strategy incorporates both standard and optional tests using different testplans. The standard testplan is shipped with each system. The testplans for the optional tests are included with the diagnostic test fixtures. The standard tests verify both the core system instrument operation and internal test paths. It uses a testplan called "U8971A_DGN.tpa", simply called "DGN".
Running Diagnostics 3 Configuring and Running the DGN Testplan 1 Start Agilent Test Exec SL by clicking this icon on the desktop: 2 Load "U8971A_DGN.tpa" testplan into TestExec SL. "DGN" Testplans are located in this directory: C:\Program Files\Agilent\TS-5000 System Software\testplan\ DGN\ 3 Before running the testplan, identify the instrumentation you want to test in the first lines of the testplan as shown on Figure 3-1.
3 Running Diagnostics Figure 3-1 Specify the Instruments to be Tested in DGN Testplan 4 Click to run the testplan. The testplan will run tests on each system instrument configuration specified in step 3 and display the results of the testing. 3-4 TS-8900 Functional Test System Diagnostics User Guide Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
Running Diagnostics 3 Configuring and Running the CEDGN Testplan 1 Start Agilent Test Exec SL by clicking this icon on the desktop: 2 There are multiple testplans for TS-8900 Functional Test Systems. The each testplan is for specific instrument, module or card that is accessible at the Express Connect interface. Table 3-1 CEDGN Testplan Description and Flow Testplan Filename Topology filename Description CEDGN_E878XA_PinMatrix.tpa CEDGN_E878XA_PinMatrix.
3 Running Diagnostics C:\Program Files\Agilent\TS-5000 System Software\Service\ U8970A\Ust 4 Load the topology file (*. ust) of the hardware to be tested and edit the Modules folder content: • Disable - Check to disable. Uncheck to enable. Depends on availability of the module in the system. • Set "Slot #" - Depends on the slot number the card is installed on the Switch Load Unit. (Applicable to SLU cards only) • Set "Instrument Descriptor/Resource" - LXI, PXI, GPIB module address.
TS-8900 Functional Test System Diagnostics User Guide Figure 3-2 Specify the Instruments to be Tested in CEDGN Testplan Running Diagnostics 3 3-7 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
3 Running Diagnostics Installing the Diagnostic Test Fixture While running the CEDGN testplan, specific diagnostic test fixtures are required to install on the Test System Interface based on the message prompted. Figure 3-3 shows the CEDGN Board #1 (U8970-66603) to TC1 and CEDGN Board #3 (U8970-66605) to TC4. Both boards are linked by Cable (E6170-61607). This fixture configuration is used to test the connectivity and instrumentations functionality in TC4.
Running Diagnostics 3 Figure 3-3 Installing CEDGN Board #1 (U8970-66603) to TC1 and CEDGN Board #3 (U8970-66605) to TC4 with both boards linked by cable (E6170-61607). TS-8900 Functional Test System Diagnostics User Guide Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
3 Running Diagnostics Figure 3-4 shows connection of 8-Ch Heavy Duty Card diagnostic cable E6170-61618. The Positronic connector on HP2 and cable to Abuses. This diagnostic cable is used to test the connectivity of 8-Ch Heavy Duty Card on HP2. 3-10 TS-8900 Functional Test System Diagnostics User Guide Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
Running Diagnostics 3 Figure 3-4 Installing 8-Ch Heavy Duty Card diagnostic cable E6170-61618 on HP2 and BNC connectors to Abus1, Abus2, Abus3 and Abus4. TS-8900 Functional Test System Diagnostics User Guide Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
3 Running Diagnostics Resolving Test Failures Test failures can be caused by improper switch settings on the Diagnostic Test Fixture, improper system equipment specification, or the actual test system instrument failure. Before troubleshooting the system equipment, make sure the test fixture switches are properly set and the test system instrumentation is properly specified.
Agilent TS-8900 Functional Test System Diagnostics User Guide 4 Diagnostic Testing Details DGN Testplan Description and Flow 4-2 Test Fixture Description 4-15 CEDGN Testplan Description and Flow 4-21 Agilent Technologies Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
4 Diagnostic Testing Details DGN Testplan Description and Flow The standard DGN diagnostic program consists of tests designed for specific system subassemblies. Generally, DGN tests are organized as follows: • System reset. • Test of other instruments, such as tests of the E6198A Switch/Load Unit , 3352xA Arbitrary Waveform Generator , 53220A Counter, power supply, and etc. • Several tests dedicated to Pin Matrix Modules - open/close relay tests.
Diagnostic Testing Details 4 Test 003: 335xA GPIB Arbitrary Waveform Generator • This tests group consists of 2 tests. First test is performing instrument self test, waits for the instrument to complete the test, and queries the instrument for the results. Second test is generating 5V from Arbitrary Waveform Generator and measure using DMM as shown in Figure 4-1 below. TS-8900 Functional Test System Diagnostics User Guide Artisan Technology Group - Quality Instrumentation ...
4 4-4 Diagnostic Testing Details TS-8900 Functional Test System Diagnostics User Guide Figure 4-1 Generate 5V from Arbitrary Waveform Generator and measure using DMM. Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
Diagnostic Testing Details 4 Test 004: 53220A GPIB Universal Frequency Counter • This test group consists of 3 tests. First test is performing instrument self test, waits for the instrument to complete the test, and queries the instrument for the results. • For second test, counter is connected to the DMM. The counter output is enabled and the expected 50 ohms resistance present in matrix path is measured using the DMM and compared to limits.
4 4-6 Diagnostic Testing Details TS-8900 Functional Test System Diagnostics User Guide Figure 4-2 Test Counter 50 Ohm and Hi Z for Channel1. Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
Diagnostic Testing Details 4 • Third test is required only if ARB is present. The counter output is connected to the ARB output and measures the output signal properties. See Figure 4-3 below. TS-8900 Functional Test System Diagnostics User Guide Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
4 4-8 Diagnostic Testing Details TS-8900 Functional Test System Diagnostics User Guide Figure 4-3 Counter measures Arbitrary Waveform Generator output signal properties. Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
Diagnostic Testing Details 4 Test 005: L4451A LXI Digital/Analog Converter • This test group performs instrument self test, waits for the instrument to complete the test, and queries the instrument for the results. Test 006: L453xA LXI Digitizer • This test group consists of 3 tests. First test performs instrument self test, waits for the instrument to complete the test, and queries the instrument for the results. • The second test is required only if ARB is present.
4 4-10 Diagnostic Testing Details TS-8900 Functional Test System Diagnostics User Guide Figure 4-4 Digitizer captures Arbitrary Waveform Generator output signal properties. Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
Diagnostic Testing Details 4 Test 007: Power Supply • This test group performs instrument self test, waits for the instrument to complete the test, and queries the instrument for the results. Repeat test for all selected Agilent power supply. Test 008: Pin Matrix Relay Open/Close • This test group consists of matrix row close and open tests. Figure 4-5 and Figure 4-6 illustrate measure short and open by DMM for matrix001:row-1 to matrix001:row-n, where n = 2, 3. . . 63, 64.
4 4-12 Diagnostic Testing Details TS-8900 Functional Test System Diagnostics User Guide Figure 4-5 Measure short by DMM for matrix row 1 and matrix row-n open. Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
TS-8900 Functional Test System Diagnostics User Guide Diagnostic Testing Details Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com 4 4-13 Figure 4-6 Measure open by DMM for matrix row 1 and matrix row-n open.
4 Diagnostic Testing Details Test 009: M9216A PXI Data Acquisition (DAQ) Test 009: M9187A PXI Digital IO Test 009: M9185A PXI Digital/Analog Converter • This test group gets information from each selected PXI modules. 4-14 TS-8900 Functional Test System Diagnostics User Guide Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
Diagnostic Testing Details 4 Test Fixture Description The test system uses Express Connect interface, test fixtures below are required to run the CEDGN testplans. Refer to Appendix A for more details. U8970-66603 PCA-DIAGNOSTIC NUMBER 1 (Board #1) Figure 4-7 CEDGN Board #1 (U8970-66603 PCA- Diagnostic Number 1) TS-8900 Functional Test System Diagnostics User Guide Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
4 Diagnostic Testing Details U8970-66604 PCA-DIAGNOSTIC NUMBER 2 (Board #2) Figure 4-8 CEDGN Board #2 (U8970-66604 PCA- Diagnostic Number 2) U8970-66605 PCA-DIAGNOSTIC NUMBER 3 (Board #3) Figure 4-9 CEDGN Board #3 (U8970-66605 PCA- Diagnostic Number 3) 4-16 TS-8900 Functional Test System Diagnostics User Guide Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
Diagnostic Testing Details 4 U8970-66606 PCA-DIAGNOSTIC NUMBER 4 (Board #4) Figure 4-10 CEDGN Board #4 (U8970-66606 PCA- Diagnostic Number 4) U8970-66607 PCA-DIAGNOSTIC NUMBER 5 (Board #5) Figure 4-11 CEDGN Board #5 (U8970-66607 PCA- Diagnostic Number 5) TS-8900 Functional Test System Diagnostics User Guide Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
4 Diagnostic Testing Details E6170-61607 CABLE 2X 2X32 IDC PIN 1 TO PIN 1 Figure 4-12 CEDGN Cable (E6170-61607 Cable 2X 2X32 IDC Pin 1 To Pin 1) E6170-61618 Diagnostic Cable Figure 4-13 Diagnostic Cable (E6170-61618) for 8 Channel Heavy Duty Card 4-18 TS-8900 Functional Test System Diagnostics User Guide Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
Diagnostic Testing Details 4 E2240-67012 CEDGN LOOPBACK CONNECTOR Figure 4-14 CEDGN Loopback Connector (E2240-67012) CEDGN test load module for E6175A, U7177A, and N9377A E6170-66520 820OHM LOADED MEZZANINE CARD Figure 4-15 820OHM Loaded Mezzanine Card (E6170-66520) CEDGN test load module for N9379A TS-8900 Functional Test System Diagnostics User Guide Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
4 Diagnostic Testing Details E6170-61619 CABLE-HEAVY DUTY LOAD JUMPER Figure 4-16 Heavy Duty Load Card Load Loopback (E6170-61619) CEDGN test load for E6178B 4-20 TS-8900 Functional Test System Diagnostics User Guide Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
Diagnostic Testing Details 4 CEDGN Testplan Description and Flow There are multiple testplans for TS-8900 Functional Test System. Each testplan tests specific instrument / module / card. Each testplan is accompanied by a topolgy file. Below is the list of CEDGN testplans. Table 4-1 CEDGN Testplan Description and Flow Testplan Filename Topology filename Description CEDGN_E878XA_PinMatrix.tpa CEDGN_E878XA_PinMatrix.ust SLU E8782A/E8783A Pin Matrix Card CEDGN Test CEDGN_E6175A_LoadCard.
4 Diagnostic Testing Details CEDGN_E878XA_PinMatrix.tpa (Pin Matrix CEDGN Test Flow) The testsplan supports testing of one E8782A and four E8783A Pin Matrix Card. User is required to set the variables to reflect the system under test. 1 = present, 0 = not present bMtrx1 = 1st Pin Matrix E8782A bMtrx2 = 2ndPin Matrix E8783A bMtrx3 = 3rd Pin Matrix E8783A bMtrx4 = 4th Pin Matrix E8783A bMtrx5 = 5th Pin Matrix E8783A Device that user sets as not present will not be tested.
Diagnostic Testing Details 4 Matrix2 E8783A Row33-Row64 Tests • Setup DMM for 4W resistance measurement. • Perform Matrix2 loopback resistance measurements between Row33-Row34, Row35-Row36, until Row63-Row64. Matrix3 Row1-Row32 • Display instruction to Install CEDGN Board #5 on appropriate TC. S1 and S2 on CEDGN Board #5 are all toggled to ON (CLOSE). Matrix3 E8783A Row1-Row32 Tests • Setup DMM for 4W resistance measurement.
4 Diagnostic Testing Details Matrix4 E8783A Row33-Row64 Tests • Setup DMM for 4W resistance measurement. • Perform Matrix4 loopback resistance measurements between Row33-Row34, Row35-Row36, until Row63-Row64. Matrix5 Row1-Row32 • Display instruction to Install CEDGN Board #5 on TC8. S1 and S2 on CEDGN Board #5 are all toggled to ON (CLOSE). Matrix5 E8783A Row1-Row32 Tests • Setup DMM for 4W resistance measurement.
Diagnostic Testing Details 4 Matrix 2 Aux 33-40 Testing • Display instruction to Install CEDGN Board #2 on TC7. S1 and S2 on CEDGN Board #2 are toggle to DOWN position. • Setup DMM for 4W resistance measurement. • Perform Matrix2 loopback resistance measurements between Aux33-Aux34, Aux35-Aux36, Aux37-Aux38 and Aux39-Aux40. Figure 4-17 Pin Matrix Loopback Test TS-8900 Functional Test System Diagnostics User Guide Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.
4 Diagnostic Testing Details CEDGN_E6175A_LoadCard.tpa (8 Channel Load Card CEDGN Test Flow) The testsplan supports testing of eight E6175A Load Cards. User is required to set the variables to reflect the system under test. 1 = present, 0 = not present bLC8CH_1 = 1st E6175A bLC8CH_2 = 2nd E6175A bLC8CH_3 = 3rd E6175A bLC8CH_4 = 4th E6175A bLC8CH_5 = 5th E6175A bLC8CH_6 = 6th E6175A bLC8CH_7 = 7th E6175A bLC8CH_8 = 8th E6175A Device that user sets as not present will not be tested.
Diagnostic Testing Details 4 E6175A 8 Channel Load Card LC8CH_2 Tests • Display instruction to Install load module E2240-67012 on LC8CH_2 Load Card. • Display instruction to install CEDGN Board #1 on TC1 and CEDGN Board #4 on appropriate TC. Install Cable E6170-61607 from SLU Utility Connector to J2 of Board #4. Install Cable E6170-61607 from J1 of Board #1 to J1 of Board #4. • Setup DMM to perform 2W resistance measurement.
4 Diagnostic Testing Details • Measure LC8CH_4 load card sense resistor path: CH1 & CH2 Current Sense, CH3 & CH4 Current Sense, CH5 & CH6 Current Sense, and CH7 & CH8 Current Sense via Row25 and Row26. E6175A 8 Channel Load Card LC8CH_5 Tests • Display instruction to Install load module E2240-67012 on LC8CH_5 Load Card. • Display instruction to install CEDGN Board #1 on TC1 and CEDGN Board #4 on appropriate TC. Install Cable E6170-61607 from SLU Utility Connector to J2 of Board #4.
Diagnostic Testing Details 4 E6175A 8 Channel Load Card LC8CH_7 Tests • Display instruction to Install load module E2240-67012 on LC8CH_7 Load Card. • Display instruction to install CEDGN Board #1 on TC1 and CEDGN Board #4 on appropriate TC. Install Cable E6170-61607 from SLU Utility Connector to J2 of Board #4. Install Cable E6170-61607 from J1 of Board #1 to J1 of Board #4. • Setup DMM to perform 2W resistance measurement.
4 Diagnostic Testing Details Figure 4-18 E6175A 8 Channel Load Card Test 4-30 TS-8900 Functional Test System Diagnostics User Guide Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
Diagnostic Testing Details 4 CEDGN_U7177A_LoadCard.tpa (24 Channel Load Card CEDGN Test Flow) The testsplan supports testing of eight U7177A Load Cards. User is required to set the variables to reflect the system under test. 1 = present, 0 = not present bLC24CH_1 = 1st U7177A bLC24CH_2 = 2nd U7177A bLC24CH_3 = 3rd U7177A bLC24CH_4 = 4th U7177A bLC24CH_5 = 5th U7177A bLC24CH_6 = 6th U7177A bLC24CH_7 = 7th U7177A bLC24CH_8 = 8th U7177A Device that user sets as not present will not be tested.
4 Diagnostic Testing Details U7177A 24 Channel Load Card LC24CH_2 Tests • Display instruction to Install load module E2240-67012 on LC24CH_2 Load Card. • Display instruction to install CEDGN Board #1 on TC1 and CEDGN Board #4 on appropriate TC. Install Cable E6170-61607 from SLU Utility Connector to J2 of Board #4. Install Cable E6170-61607 from J1 of Board #1 to J1 of Board #4. • Setup DMM to perform 2W resistance measurement.
Diagnostic Testing Details 4 U7177A 24 Channel Load Card LC24CH_4 Tests • Display instruction to Install load module E2240-67012 on LC24CH_4 Load Card. • Display instruction to install CEDGN Board #1 on TC1 and CEDGN Board #4 on appropriate TC. Install Cable E6170-61607 from SLU Utility Connector to J2 of Board #4. Install Cable E6170-61607 from J1 of Board #1 to J1 of Board #4. • Setup DMM to perform 2W resistance measurement.
4 Diagnostic Testing Details U7177A 24 Channel Load Card LC24CH_6 Tests • Display instruction to Install load module E2240-67012 on LC24CH_6 Load Card. • Display instruction to install CEDGN Board #1 on TC1 and CEDGN Board #4 on appropriate TC. Install Cable E6170-61607 from SLU Utility Connector to J2 of Board #4. Install Cable E6170-61607 from J1 of Board #1 to J1 of Board #4. • Setup DMM to perform 2W resistance measurement.
Diagnostic Testing Details 4 U7177A 24 Channel Load Card LC24CH_8 Tests • Display instruction to Install load module E2240-67012 on LC24CH_8 Load Card. • Display instruction to install CEDGN Board #1 on TC1 and CEDGN Board #4 on appropriate TC. Install Cable E6170-61607 from SLU Utility Connector to J2 of Board #4. Install Cable E6170-61607 from J1 of Board #1 to J1 of Board #4. • Setup DMM to perform 2W resistance measurement.
4 Diagnostic Testing Details CEDGN_E6178B_LoadCard.tpa (8 Channel High Power Load Card CEDGN Test Flow) The testsplan supports testing of eight E6178B High Power Load Cards. User is required to set the variables to reflect the system under test. 1 = present, 0 = not present bLC8CH_1 = 1st E6178B bLC8CH_2 = 2nd E6178B bLC8CH_3 = 3rd E6178B bLC8CH_4 = 4th E6178B bLC8CH_5 = 5th E6178B bLC8CH_6 = 6th E6178B bLC8CH_7 = 7th E6178B bLC8CH_8 = 8th E6178B Device that user sets as not present will not be tested.
Diagnostic Testing Details 4 E6178B 8 Channel High Power Load Card LC8CH_1 Tests • Display instruction to disconnect power supply from (P100)PWR1 and (P200)PWR2 of E6178B_1. Short PWR1 to PWR2 of E6178B_1. Install load loopback cable E6170-61619 on E6178B_1. Install Diagnostic Cable (E6170-61618) on appropriate HP connector. • Setup DMM to perform 2W resistance measurement.
4 Diagnostic Testing Details E6178B 8 Channel High Power Load Card LC8CH_3 Tests • Display instruction to disconnect power supply from (P100)PWR1 and (P200)PWR2 of E6178B_3. Short PWR1 to PWR2 of E6178B_3. Install load loopback cable E6170-61619 on E6178B_3. Install Diagnostic Cable (E6170-61618) on appropriate HP connector. • Setup DMM to perform 2W resistance measurement.
Diagnostic Testing Details 4 E6178B 8 Channel High Power Load Card LC8CH_5 Tests • Display instruction to disconnect power supply from (P100)PWR1 and (P200)PWR2 of E6178B_5. Short PWR1 to PWR2 of E6178B_5. Install load loopback cable E6170-61619 on E6178B_5. Install Diagnostic Cable (E6170-61618) on appropriate HP connector. • Setup DMM to perform 2W resistance measurement.
4 Diagnostic Testing Details E6178B 8 Channel High Power Load Card LC8CH_7 Tests • Display instruction to disconnect power supply from (P100)PWR1 and (P200)PWR2 of E6178B_7. Short PWR1 to PWR2 of E6178B_7. Install load loopback cable E6170-61619 on E6178B_7. Install Diagnostic Cable (E6170-61618) on appropriate HP connector. • Setup DMM to perform 2W resistance measurement.
Diagnostic Testing Details 4 Clean Up Information • Display instruction to remove all Short from PWR1 and PWR2 of all E6178B load cards. Figure 4-20 E6178B 8 Channel High Power Load Card Test TS-8900 Functional Test System Diagnostics User Guide Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
4 Diagnostic Testing Details CEDGN_N9377A_LoadCard.tpa (16 Channel Load Card CEDGN Test Flow) The testsplan supports testing of eight N9377A Load Cards. User is required to set the variables to reflect the system under test. 1 = present, 0 = not present bLC16CH_1 = 1st N9377A bLC16CH_2 = 2nd N9377A bLC16CH_3 = 3rd N9377A bLC16CH_4 = 4th N9377A bLC16CH_5 = 5th N9377A bLC16CH_6 = 6th N9377A bLC16CH_7 = 7th N9377A bLC16CH_8 = 8th N9377A Device that user sets as not present will not be tested.
Diagnostic Testing Details 4 N9377A 16 Channel Load Card LC16CH_1 Tests • Display instruction to Install load module E2240-67012 on LC16CH_1 Load Card. • For LC16CH_1 on TC1, display instruction to install CEDGN Board #4 on TC1 and Cable E6170-61607 from SLU Utility Connector to J2 of Board #4. • For LC16CH_1 on other TC, display instruction to install CEDGN Board #1 on TC1 and CEDGN Board #4 on appropriate TC. Install Cable E6170-61607 from SLU Utility Connector to J2 of Board #4.
4 Diagnostic Testing Details N9377A 16 Channel Load Card LC16CH_3 Tests • Display instruction to Install load module E2240-67012 on LC16CH_3 Load Card. • Display instruction to install CEDGN Board #1 on TC1 and CEDGN Board #4 on appropriate TC. Install Cable E6170-61607 from SLU Utility Connector to J2 of Board #4. Install Cable E6170-61607 from J1 of Board #1 to J1 of Board #4. • Setup DMM to perform 2W resistance measurement. • Measure LC16CH_3 load card loopback path resistance from CH1-CH2 via Loadx.
Diagnostic Testing Details 4 N9377A 16 Channel Load Card LC16CH_5 Tests • Display instruction to Install load module E2240-67012 on LC16CH_5 Load Card. • Display instruction to install CEDGN Board #1 on TC1 and CEDGN Board #4 on appropriate TC. Install Cable E6170-61607 from SLU Utility Connector to J2 of Board #4. Install Cable E6170-61607 from J1 of Board #1 to J1 of Board #4. • Setup DMM to perform 2W resistance measurement. • Measure LC16CH_5 load card loopback path resistance from CH1-CH2 via Loadx.
4 Diagnostic Testing Details N9377A 16 Channel Load Card LC16CH_7 Tests • Display instruction to Install load module E2240-67012 on LC16CH_7 Load Card. • Display instruction to install CEDGN Board #1 on TC1 and CEDGN Board #4 on appropriate TC. Install Cable E6170-61607 from SLU Utility Connector to J2 of Board #4. Install Cable E6170-61607 from J1 of Board #1 to J1 of Board #4. • Setup DMM to perform 2W resistance measurement. • Measure LC16CH_7 load card loopback path resistance from CH1-CH2 via Loadx.
Diagnostic Testing Details 4 Figure 4-21 N9377A 16 Channel Load Card Test TS-8900 Functional Test System Diagnostics User Guide Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
4 Diagnostic Testing Details CEDGN_N9379A_LoadCard.tpa (48 Channel Load Card CEDGN Test Flow) The testsplan supports testing of four N9379A Load Cards. User is required to set the variables to reflect the system under test. 1 = present, 0 = not present bLC48CH_1 = 1st N9379A bLC48CH_2 = 2nd N9379A bLC48CH_3 = 3rd N9379A bLC48CH_4 = 4th N9379A NOTE: All N9379A need to be installed with load modules E6170-66520 Device that user sets as not present will not be tested.
Diagnostic Testing Details 4 N9379A 48 Channel Load Card LC48CH_2 on TC3 (CH1 until CH24) Tests • Display instruction to install CEDGN Board #1 on TC1 and CEDGN Board #4 on appropriate TC3. Install Cable E6170-61607 from J1 of Board #1 to J1 of Board #4 • Setup DMM to perform 2W resistance measurement • Measure LC48CH_2 load card loopback path resistance from CH1-Com1 via Load1.1/Load1.2 until CH12-Com12 via Load12.1/Load12.2 using Row 1 and Row8. CH13-Com13 via Load13.1/Load13.
4 Diagnostic Testing Details N9379A 48 Channel Load Card LC48CH_3 on TC6 (CH25 until CH48) Tests • Display instruction to install CEDGN Board #1 on TC1 and CEDGN Board #4 on appropriate TC6. Install Cable E6170-61607 from J1 of Board #1 to J1 of Board #4. • Setup DMM to perform 2W resistance measurement. • Measure LC48CH_3 load card loopback path resistance from CH25-Com25 via Load25.1/Load25.2 until CH36-Com36 via Load36.1/Load36.2 using Row 1 and Row8. CH37-Com37 via Load37.1/Load37.
Diagnostic Testing Details 4 Figure 4-22 N9379A 48 Channel Load Card Test TS-8900 Functional Test System Diagnostics User Guide Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
4 Diagnostic Testing Details CEDGN_E6198B_Utility.tpa (Switch Load Unit Utility CEDGN Test Flow) The testsplan performs testing on the SLU Utility Connector at the Express Connect front panel. SLU Utility Setup Information • Display instruction to install CEDGN Board #1 on TC1. Install Cable E6170-61607 on SLU Utility Connector to J2 of Board #1. SLU Utility System_Gnd Tests • Setup DMM to measure voltage.
Diagnostic Testing Details 4 SLU Utility FixtureID to OpenDrainOut Tests • Uses 8bit OpenDrainOut as output and 8bit FixtureID as input. • Tests OpenDrainOut and FixtureID Write/Read function covering 8bit data pattern: 10101010 (170) 01010101 (85) 11111111 (255) 00000000 (0) SLU Utility Digital In to Digital Out Tests • Uses 8bit Digital Out as output and 8bit Digital In as input.
4 Diagnostic Testing Details Figure 4-23 E6198B Switch Load Unit Utility Test 4-54 TS-8900 Functional Test System Diagnostics User Guide Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
Diagnostic Testing Details 4 CEDGN_L4451A_LXIDAC.tpa (4 Channel LXI DAC CEDGN Test Flow) The testsplan supports testing of two L4451A DAC. User is required to set the variables to reflect the system under test. 1 = present, 0 = not present bLXIDAC1 = 1st L4451A bLXIDAC2 = 2nd L4451A Device that user sets as not present will not be tested. The test starts by performing: Not Acceptable Combination Checking • Checks user settings for any not acceptable combination.
4 Diagnostic Testing Details L4451A LXI DAC2 Tests • Display instruction to install CEDGN Board #1 on TC1 and CEDGN Board #3 on appropriate TC. Install Cable E6170-61607 from J1 of Board #1 to J1 of Board #3. • Setup DMM to perform voltage measurement.
Diagnostic Testing Details 4 Figure 4-24 L4451A LXI DAC Test TS-8900 Functional Test System Diagnostics User Guide Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
4 Diagnostic Testing Details CEDGN_M9185A_DAC.tpa (8/16 Channel PXI DAC CEDGN Test Flow) The testsplan is for testing of M9185A 8/16 Channel DAC. It supports two 16 Channel DAC or three 8 Channel DAC or one 16 Channel and one 8 Channel DAC. User is required to set the variables to reflect the system under test.
Diagnostic Testing Details 4 M9185A 16 Channel PXI DAC1 (CH1-CH8) on TC1 Tests • Display instruction to install CEDGN Board #2 on TC1 and toggle S1 and S2 on Board #2 to DOWN position. • Setup DMM to perform voltage measurement.
4 Diagnostic Testing Details CH1 with/without sense CH2 with/without sense CH3 with/without sense CH4 with/without sense CH5 with/without sense CH6 with/without sense CH7 with/without sense CH8 with/without sense • Display instruction to toggle S1 and S2 on Board #2 to UP position. • Setup PXI DAC16_2 to output 16V on CH8 when External Trigger is detected. • Use SLU internal DAC1 to generate a simulated trigger to PXI DAC16_2. • Measure for 16V at PXI DAC16_2 CH8 output for trigger response.
Diagnostic Testing Details 4 CH5 with/without sense CH6 with/without sense CH7 with/without sense CH8 with/without sense • Display instruction to toggle S1 and S2 on Board #2 to UP position. • Setup PXI DAC8_1 to output 16V on CH8 when External Trigger is detected. • Use SLU internal DAC1 to generate a simulated trigger to PXI DAC8_1. • Measure for 16V at PXI DAC8_1 CH8 output for trigger response. M9185A 8 Channel PXI DAC2 (CH1-CH8) Tests • Display instruction to install CEDGN Board #1 on TC1.
4 Diagnostic Testing Details • Use SLU internal DAC1 to generate a simulated trigger to PXI DAC8_2. • Measure for 16V at PXI DAC8_2 CH8 output for trigger response. M9185A 8 Channel PXI DAC3 (CH1-CH8) Tests • Display instruction to install CEDGN Board #1 on TC1. Install CEDGN Board #2 on appropriate TC and toggle S1 and S2 on Board #2 to DOWN position. Install Cable E6170-61607 on J1 of Board #1 to J1 of Board #2. • Setup DMM to perform voltage measurement.
Diagnostic Testing Details 4 Figure 4-25 M9185A PXI DAC Output Test TS-8900 Functional Test System Diagnostics User Guide Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
4 Diagnostic Testing Details Figure 4-26 M9185A PXI DAC Trigger Test 4-64 TS-8900 Functional Test System Diagnostics User Guide Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
Diagnostic Testing Details 4 CEDGN_M9186A_VI.tpa (VI Source CEDGN Test Flow) The testsplan supports testing of two M9186A VI Source modules. It is used to verify the connectivity of VI Source module Safety Interlock on the Express Connect TC connector. Test will abort with EXCEPTION ENCOUNTERED error if safety interlock is not installed. User is required to set the variables to reflect the system under test.
4 Diagnostic Testing Details M9186A VI Source Module VI2 Tests • Display instruction to install CEDGN Board #3 on appropriate TC.
Diagnostic Testing Details 4 CEDGN_M9187A_DIO.tpa (DIO CEDGN Test Flow) The testsplan supports testing of two M9187A DIO modules. User is required to set the variables to reflect the system under test. 1 = present, 0 = not present bDIO1 = 1st M9187A DIO bDIO2 = 2nd M9187A DIO Device that user sets as not present will not be tested. The test starts by performing: Not Acceptable Combination Checking • Checks user settings for any not acceptable combination.
4 Diagnostic Testing Details M9187A DIO1 on TC4 (Channel 17-32) Tests • Setup SLU internal DAC1 to output 5V for use on DIO1 Vext. • Display instruction to install CEDGN Board #1 on TC1 and CEDGN Board #3 on TC4. Install Cable E6170-61607 on J1 of Board #1 to J1 of Board #3.
Diagnostic Testing Details 4 Figure 4-27 M9187A DIO Test TS-8900 Functional Test System Diagnostics User Guide Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
4 Diagnostic Testing Details CEDGN_M9216A_DAQ.tpa (High Voltage Data Acquisition CEDGN Test Flow) The testsplan supports testing of four M9216A DAQ. User is required to set the variables to reflect the system under test. 1 = present, 0 = not present bDAQ1 = 1st M9216A bDAQ2 = 2nd M9216A bDAQ3 = 3rd M9216A bDAQ4 = 4th M9216A Device that user sets as not present will not be tested. The test starts by performing: Not Acceptable Combination Checking • Checks user settings for any not acceptable combination.
TS-8900 Functional Test System Diagnostics User Guide M9216A DAQ1 Tests • Display instruction to install CEDGN Board #1 on TC1 and CEDGN Board #5 on appropriate TC. OPEN all 16 actuators on switch S1 and S2 of Board #5. Install Cable E6170-61607 from J1 of Board #1 to J1 of Board #5. • Setup SLU internal DAC1 to output 5V. • Route 5V from DAC1 to individual DAQ channel and use DAQ to measure each channel.DAC1 EarthGnd is connected to DAQ ground via UUTCommon of PinMatrix.
4 Diagnostic Testing Details DAQ1 7B DAQ1 7C DAQ1 7D DAQ1 8A DAQ1 8B DAQ1 8C DAQ1 8D (via Row26) (via Row27) (via Row28) (via Row29) (via Row30) (via Row31) (via Row32) M9216A DAQ2 Tests • Display instruction to install CEDGN Board #1 on TC1 and CEDGN Board #5 on appropriate TC. OPEN all 16 actuators on switch S1 and S2 of Board #5. Install Cable E6170-61607 from J1 of Board #1 to J1 of Board #5. • Setup SLU internal DAC1 to output 5V.
Diagnostic Testing Details DAQ2 8A DAQ2 8B DAQ2 8C DAQ2 8D 4 (via Row29) (via Row30) (via Row31) (via Row32) M9216A DAQ3 Tests • Display instruction to install CEDGN Board #1 on TC1 and CEDGN Board #5 on appropriate TC. OPEN all 16 actuators on switch S1 and S2 of Board #5. Install Cable E6170-61607 from J1 of Board #1 to J1 of Board #5. • Setup SLU internal DAC1 to output 5V. • Route 5V from DAC1 to individual DAQ channel and use DAQ to measure each channel.
4 Diagnostic Testing Details M9216A DAQ4 Tests • Display instruction to install CEDGN Board #1 on TC1 and CEDGN Board #5 on appropriate TC. OPEN all 16 actuators on switch S1 and S2 of Board #5. Install Cable E6170-61607 from J1 of Board #1 to J1 of Board #5. • Setup SLU internal DAC1 to output 5V. • Route 5V from DAC1 to individual DAQ channel and use DAQ to measure each channel.DAC1 EarthGnd is connected to DAQ ground via UUTCommon of PinMatrix.
Diagnostic Testing Details 4 Figure 4-28 M9216A High Voltage Data Acquisition Test TS-8900 Functional Test System Diagnostics User Guide Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
4 Diagnostic Testing Details CEDGN_RS232.tpa (RS232 Serial COM CEDGN Test Flow) The testsplan supports testing of four RS232 COM ports. It is a loopback test where Tx and Rx of the COM port is connected together. User is required to set the variables to reflect the system under test. 1 = present, 0 = not present bCOM1 = 1st COM Port bCOM2 = 2nd COM Port bCOM3 = 3rd COM Port bCOM4 = 4th COM Port Device that user sets as not present will not be tested.
Diagnostic Testing Details 4 COM4 on TC7 Tests • Display instruction to Install CEDGN Board #1 on TC7. • Use COM4 to transmit characters “This is test com4”. • Use COM4 to receive and compare to ensure the received characters matches the transmitted characters. TS-8900 Functional Test System Diagnostics User Guide Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
4 Diagnostic Testing Details CEDGN_SoftingCAN.tpa (Softing CAN CEDGN Test Flow) The testsplan supports testing of two Softing CAN cards. Each card has two CAN ports. User is required to set the variables to reflect the system under test. 1 = present, 0 = not present bCAN1_1_AND_CAN1_2_TC1 = 1st Softing CAN card with CAN ports on TC1 bCAN2_1_TC3 = 2nd Softing CAN card with CAN port 1 on TC3 bCAN2_2_TC5 = 2nd Softing CAN card with CAN port 2 on TC5 Device that user sets as not present will not be tested.
Agilent TS-8900 Functional Test System Diagnostics User Guide A Diagnostic Test Fixture Details Celsior Test Fixture Schematics A-2 Agilent Technologies Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
Figure A-1 U8970-66603 PCA-Diagnostic Number 1 (Board #1) TS-8900 Functional Test System Diagnostics User Guide Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
TS-8900 Functional Test System Diagnostics User Guide Schematic for U8970-66604 PCA-Diagnostic Number 2 (Board #2) Figure A-2 U8970-66604 PCA-Diagnostic Number 2 (Board #2) Diagnostic Test Fixture Details A A-3 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
A Figure A-3 U8970-66605 PCA-Diagnostic Number 3 (Board #3) Diagnostic Test Fixture Details A-4 Schematic for U8970-66605 PCA-Diagnostic Number 3 (Board #3) TS-8900 Functional Test System Diagnostics User Guide Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
TS-8900 Functional Test System Diagnostics User Guide Schematic for U8970-66606 PCA-Diagnostic Number 4 (Board #4) Figure A-4 U8970-66606 PCA-Diagnostic Number 4 (Board #4) Diagnostic Test Fixture Details A A-5 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
A Figure A-5 U8970-66607 PCA-Diagnostic Number 5 (Board #5) Diagnostic Test Fixture Details A-6 Schematic for U8970-66607 PCA-Diagnostic Number 5 (Board #5) TS-8900 Functional Test System Diagnostics User Guide Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
Diagnostic Test Fixture Details A Table A-1 CEDGN Board #2 switch mapping summary matrix row # switch TC connector pin row 2 S1 (UP position) PXI DAC DMM_H and PXI DAC DMM_C S1 (DOWN position) PXI DAC1H S1 (UP position) PXI DAC DMM_L S1 (DOWN position) PXI DAC1L S2 (UP position) PXI DAC TRIGGER S2 (DOWN position) PXI DAC2H S2 (UP position) PXI DAC ZGNDF S2 (DOWN position) PXI DAC2L row 4 row 6 row 8 TS-8900 Functional Test System Diagnostics User Guide Artisan Technology Group - Qua
A Diagnostic Test Fixture Details Table A-2 CEDGN Board #5 switch mapping summary matrix row # switch TC connector pin row 1 S1 Actuator 1 (ON) -CLOSE row 2 S1 Actuator 1 (OFF) -OPEN NC S1 Actuator 2 (ON) -CLOSE row 4 S1 Actuator 2(OFF) -OPEN NC S1 Actuator 3 (ON) -CLOSE row 6 S1 Actuator 3 (OFF) -OPEN NC S1 Actuator 4 (ON) -CLOSE row 8 S1 Actuator 4 (OFF) -OPEN NC S1 Actuator 5 (ON) -CLOSE row 10 S1 Actuator 5 (OFF) -OPEN NC S1 Actuator 6 (ON) -CLOSE row 12 S1 Actuator 6 (OFF) -
TS-8900 Functional Test System Diagnostics User Guide Schematic for E6170-61618 Diagnostic Cable for 8 Channel Heavy Duty Card Figure A-6 E6170-61618 Diagnostic Cable for 8 Channel Heavy Duty Card Diagnostic Test Fixture Details A A-9 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
A Figure A-7 E6170-61607 Cable (Cable -2X 2X32 IDC PIN1 to PIN1) Diagnostic Test Fixture Details A-10 Schematic for E6170-61607 Cable (Cable -2X 2X32 IDC PIN1 to PIN1) TS-8900 Functional Test System Diagnostics User Guide Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.
Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment • FAST SHIPPING AND DELIVERY • TENS OF THOUSANDS OF IN-STOCK ITEMS • EQUIPMENT DEMOS • HUNDREDS OF MANUFACTURERS SUPPORTED • LEASING/MONTHLY RENTALS • ITAR CERTIFIED SECURE ASSET SOLUTIONS SERVICE CENTER REPAIRS Experienced engineers and technicians on staff at our full-service, in-house repair center WE BUY USED EQUIPMENT Sell your excess, underutilized, and idle used equipment We also offer credit for buy-back