Specifications
80 Using the DMM and ARB Chapter 4
Figure 4-11. Typical Segment Number Value
Figure 4-12. Typical Segment Number Value
20 segments in
20 segments in
Second Sequence
First Sequence
Waveform time for both sequences is: total of (segments * clock time for each sequence) =
Since both sequence time is the same, the total time = 2 * 0.002 = 0.004 = 250 Hz
Dwell count of
49 for 20 segments
in First Sequence
Dwell count of
49 for 20 segments
in Second Sequence
20 * ((49 * 0.000002) + 0.000002) = 20 * (0.000098 + 0.000002) = 20 * 0.0001 = 0.002 per sequence
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com