Specifications

Planning System Testplans 53Chapter 3
Instruments & Measurements: The instrument section is used to describe
what instruments are needed for the test and how they should be
programmed.
Connections: In the connections section, describe how the resource will be
connected (e.g., through a 32-pin Matrix module, directly, or indirectly
through the Matrix module auxiliary channels connector). The 32-Pin
Matrix module has 16 external instrument connections which can be
connected to four analog buses (AB1 to AB4) or the UUT common ground
bus. The Digital Multimeter, one channel of the Arbitrary Waveform
Generator, one or more channels of the DAC, the Counter, the Digitizer, and
the Measurement Control module resources are generally connected to the
UUT through Matrix modules. Loads in the load box are typically connected
to the UUT through the TC1 - TC8 connectors (Test System Interface) or the
one-one-two expander blocks (Mass Interconnect).
The 32-Pin Matrix modules can handle signals up to 200 volts peak. Other
instruments may have even lower voltage restrictions.
Caution Input signals greater than 100 V
peak/dc
voltages can damage the
digitizer. Input signals greater than 200 V
peak/dc
voltages can
damage the 32-Pin Matrix modules. Other instruments may
have lower voltage restrictions. Voltages exceeding these limits
must be attenuated before being applied to these modules to
avoid damage.
Cleanup: Cleanup describes the state the system and UUT should go to
upon completion of the test