Specifications

13
16850 Series Instrument Characteristics (continued)
16850 Series PC characteristics
Operating system Microsoft Windows
®
7 Embedded (64-bit)
Processor Core 2 Duo, M890, 3.0 GHz microprocessor
Chipset Intel Q45
System memory 4 GB
Removable Hard disk drive 500 GB
Installed on hard drive Operating system, latest revision of the logic and protocol application software, optional
application software ordered with the logic analyzer
16850 Series instrument controls
LCD touch-screen display Large 38.1-cm (15-in.) touch-screen display makes is easy to view a large number of waveforms
or states
Front-panel hot keys Dedicated hot keys for selecting run mode and disabling touch screen
Front-panel knob General-purpose knob adjusts viewing and measurement parameters
Keyboard and mouse PS/2 keyboard and USB mouse
16850 Series video display modes
Touch-screen display standard Size 38.1 cm (15 in.) diagonal
Resolution 1024 x 768
External display Simultaneous display capability Front panel and external display can be used
simultaneously at 1024 x 768 resolution
Supports up to four external monitors at up to 1600 x 1200 (with PCI video card)
Programmability
You can write programs to control the logic analyzer application from remote computers on the local area network using
COM or ASCII.
The COM automation server is part of the logic analyzer application. This software allows you to write programs to control
the logic analyzer. All measurement functionality is controllable via the COM interface.
The B4608A Remote Programming Interface (RPI) lets you remotely control a 16850 Series logic analyzer by issuing ASCII
commands to the TCP socket on port 6500. This interface is designed to be as similar as possible to the RPI on 16700
Series logic analysis systems, so that you can reuse existing programs.
The remote programming interface works through the COM automation objects, methods, and properties provided for
controlling the logic analyzer application. RPI commands are implemented as Visual Basic modules that execute COM
automation commands, translate their results, and return proper values for the RPI. You can use the B4606A advanced
customization environment to customize and add RPI commands.
Figure 8. 16850 Series programming
overview.
Computer with
Windows XP or
Windows 7
Logic analyzer
(16850/16800/16900)
or PC running Agilent
logic analyzer application
LAN connection
Distributed COM
Instrument COM
automation server