Product specifications
34
Figure 25b. Measurements after calibration with pulse modulator off
Figure 25c. Corrected pulsed-RF S-parameters and absolute power measurements
using narrowband detection
Note that the reference receiver measures the pulsed stimulus power at the set
power minus 20*log (receiver gate duty cycle) with some additional path loss
(Figure 25a). Once the reference receiver is calibrated it reads close to set power
at -10 dBm (Figure 25b). Then the pulsed stimulus level is well maintained after
turning on the pulse modulation and receiver leveling (Figure 25c).