Product specifications
13
Internal pulse generators start generating the pulses approximately 60 to
100 ns after the pulse trigger inputs at the PULSE SYNC IN – (denoted as
pulse trigger delay). The jitter of this delay is the minimum time resolution
of the system. The exact pulse trigger delay can be measured with an
oscilloscope between PULSE SYNC IN and one of the PULSE OUT (P1
through P4). If the internal clock is used to trigger the pulse generators,
this pulse trigger delay is very small and therefore it can be neglected. One
of the P1 through P4 internal pulse generators drives the internal pulse
modulators to generate pulsed-RF signals. The actual pulsed-RF signals
have a delay of approximately 30 ns from the P1 through P4 modulation
pulse at the test port with a 48 inch RF test port cable – (denoted as pulse
modulation delay). The pulse modulation delay with RF carrier frequencies
of 3.2 GHz and below is about 40 ns or larger (nearly 100 ns delay at the
low-end frequency of the analyzer). The combiner path on port 1 adds an
additional 5 ns delay compared to the thru path. The pulse modulators
switch the RF sources on and off with approximately
4 ns rise time and 10 ns fall time.
P0 pulse generator is also triggered by PULSE SYNC IN (or internal
trigger) and generates a data acquisition pulse with the same amount of
pulse trigger delay as other internal pulse generators (or zero if triggered
internally). Although the data acquisition process starts immediately with
the P0 pulse, there is approximately 250 ns data-processing delay, the
time it takes for sampled data with pulse on to become available in the
buffer. The user-specified measurement delay (the delay for P0) must take
into account these data-acquisition and pulse-modulation delays to align
the pulsed stimulus and data-acquisition window. Approximately 300 ns
for the measurement delay accounts for pulse-modulation delay and data-
acquisition delay, as well as pulse-settling time. Additional delay may be
necessary depending on the frequency, the PNA-X’s internal path switches,
and external cables and devices.
Setting up pulsed-RF measurements
The basis of pulsed-RF point-in-pulse measurements with wideband detection
method is to synchronize pulsed stimulus and data acquisition so that all
receivers measure responses only within the RF pulses. The following three
steps must be completed for successful measurements.
Step 1 Setup pulse generators and modulators
Specify pulse width, delay and period for internal pulse generators with Pulse
Generator Setup dialog (Figure 10a). Note that Pulse0 width is determined
by the IF bandwidth chosen (in step 3) thus it is not editable on the pulse
generator dialog. When internal pulse modulators are used, specify the drive
source (typically one of internal pulse generators). It is also required to disable
automatic level control (ALC) on the source port with pulse modulation
enabled. Change the leveling mode from “Internal” to “Open loop” on Power
and attenuator dialog. Otherwise the ALC will try to level the source with the
detected power level with pulse on and off, causing a source unleveled error.