User`s guide

E5810B User’s Guide 25
Installing the E5810B 2
Configuring the E5810B on a Local Network
Crossover cable connections
Figure 2-5 Connecting the E5810B to a local network via a
crossover cable
Turn on the E5810B and the IP address will be displayed on
the front panel display.
1
69
5
1
9
5
6
1
13
GPIB
RS232
+12VDC
2.0A
LAN
ICES/NMB-001
ISM GRP 1-A
N10149
KCC-REM-ATi-
WCIOP
E5810B
PC
GPIB instrument
RS-232 instrument
USB instrument
Crossover
cable
NOTE
The E5810B may require up to 75 seconds to display the default static
IP address (169.254.58.10) as it will try to find a DHCP server until a
timeout value is reached.
If you try to simultaneously attach multiple E5810Bs without changing
each unit’s default static IP address, some E5810Bs may not be able to
start operation. The E5810B detects duplicated IP addresses and will
not start operation if the IP address is already in use.
To prevent this, add a E5810B at a time and configure the E5810B to a
unique IP address before adding the additional E5810Bs.