User`s guide

E5810B User’s Guide 97
Troubleshooting the E5810B 4
Checking PC Client Connection
Table 4- 5 shows the typical Syslog messages and provides a
description of each message, where the value is in ( ), such
as (0:00:00:18). The value is the time since power- on
expressed in Day (DD):Hour (HH):Minutes (MM):Seconds
(SS).
ERROR: Firmware Update: Invalid file or directory Could not find file. Filename or directory is wrong.
ERROR: Firmware Update: Internal error Out of system resources. Reboot the E5810B and try again.
Tab le 4-4 Common Syslog messages (continued)
Message Description
Tab le 4-5 Typical Syslog messages
Message Description
(0:00:00:00) Selftest Passed Power-on self-test results.
(00:00:00.00) GPIB initialized with name : [gpib0]
(00:00:00.00) RS232 initialized with name : [COM1]
(00:00:00.00) USB initialized with name : [usb0]
GPIB, RS-232, and USB initialized.
(0:00:00:18) client: 156.140.104.184, id: 1 opened
(0:00:00:18) client: 156.140.104.184, id: 1 closed
(0:00:00:18) client: 156.140.104.184, id: 1 opened
(0:00:00:19) client: 156.140.104.184, id: 2 opened
(0:00:00:19) client: 156.140.104.184, id: 3 opened
Typical open or close messages (logged each time a client opens
or closes a connection).
(0:00:39:21) Max connections exceeded, client refused Indicates the server had the maximum number of connections
when another connection was requested, so the request