User`s guide

5-54
Agilent Technologies E5500 Phase Noise Measurement System
Expanding Your Measurement Experience
Testing the Agilent/HP 8644B Internal/External 10 MHz
Table 5-8 Parameter Data for the Agilent/HP 8644B 10 MHz Measurement
Step Parameters Data
1 Type and Range Tab
Measurement Type
Start Frequency
Stop Frequency
Minimum Number of Averages
FFT Quality
Absolute Phase Noise (using a phase locked loop)
•10Hz
•2E+6Hz
1
•4
•Fast
2 Sources Tab
Carrier Source
Frequency
Power
Carrier Source Output is
connected to:
Detector Input
Frequency
Reference Source
Frequency
Reference Source Power
VCO Tuning Parameters
Nominal Tune Constant
Tune Range +/-
Center Voltage
Input Resistance
•10E+6Hz
•7dBm
Test Set
•10E+6Hz
10 E +6 Hz (same as Carrier Source Frequency)
•16dBm
•1E+3Hz/V
•+/-10Volts
•0Volts
•600ohms
3CalTab
Phase Detector Constant
VCO Tune Constant
Phase Lock Loop Suppression
If Limit is exceeded
Measure Phase Detector Constant
Calculate from expected VCO Tune Constant
Verify calculated phase locked loop suppression
Show Suppression Graph
4 Block Diagram Tab
Carrier Source
Downconverter
Reference Source
Timebase
Phase Detector
Test Set Tune Voltage
Destination
VCO Tune Mode
Manual
•None
Agilent/HP 8644B
•None
Automatic Detector Selection
Reference Source
DCFM