User`s guide

5-32
Agilent Technologies E5500 Phase Noise Measurement System
Expanding Your Measurement Experience
Testing the Agilent/HP 8663A Internal/External 10 MHz
5 Test Set Tab
Input Attenuation
LNA Low Pass Filter
•LNAGain
•DCBlock
PLL Integrator Attenuation
•0dB
20 MHz (Auto checked)
Auto Gain (Minimum Auto Gain - 14 dB)
Not checked
•0dBm
6 Dowconverter Tab The downconverter parameters do not apply to this
measurement example.
7GraphTab
•Title
Graph Type
X Scale Minimum
X Scale Maximum
Y Scale Minimum
Y Scale Maximum
Normalize trace data to a:
Scale trace data to a new
carrier frequency of:
Shift trace data DOWN by:
Trace Smoothing Amount
Power present at input of DUT
Confidence Test using Agilent/HP 8663A Int vs Ext 10
MHz
Single-sideband Noise (dBc/Hz)
•10Hz
•4E+6Hz
•0dBc/Hz
- 170 dBc/Hz
1 Hz bandwidth
1 times the current carrier frequency
•0dB
•0
•0dB
1. The Stop Frequency depends on the analyzers configured in your phase noise system.
Table 5-4 Parameter Data for the Agilent/HP 8663A 10 MHz Measurement
Step Parameters Data