User`s guide

5-14
Agilent Technologies E5500 Phase Noise Measurement System
Expanding Your Measurement Experience
Testing the Agilent/HP 8663A Internal/External 10 MHz
Selecting Loop
Suppression
Verification
1. From the Define menu, choose Measurement; then choose the Cal tab
from the Define Measurement window.
2. In the Cal dialog box, check Verify calculated phase locked loop
supression and Always Show Suppression Graph. Select If limit is
exceeded: Show Loop Suppression Graph.
3. When you have completed these operations, click the Close button
.
Setup Considerations
for the
Agilent/HP 8663A
10 MHz Measurement
Measurement Noise Floor
The signal amplitude at the R input (Signal Input) port on the
Agilent/HP 70420A sets the measurement noise floor level. Use the
following graph to determine the amplitude required to provide a noise floor
level that is below the expected noise floor of your UUT. For more
information about this graph, refer to Chapter 18, “Reference Graphs and
Tables.