User`s guide
Agilent Technologies E5500 Phase Noise Measurement System
13-11
AM Noise Measurement Examples
AM Noise using an Agilent/HP 70420A Option 001
4 Block Diagram Tab
• Source
•AMDetector
• Down Converter
• Manual
• TestSet AM Detector
•None
5 Test Set Tab
• Input Attenuation
• LNA Low Pass Filter
•LNAGain
• Detector Maximum Input
Levels
Microwave Phase Detector
RF Phase Detector
AM Detector
• Ignore out-of-lock conditions
• Pulsed Carrier
•DCBlock
• Analyzer View
• PLL Integrator Attenuation
• Auto checked
• Auto checked
•AutoGain
•0dBm
•0dBm
•0dBm
• Not checked
• Not checked
• Not checked
• Baseband
•0.00dBm
6 Dowconverter Tab • Does not apply to this measurement example.
7GraphTab
•Title
• Graph Type
• X Scale Minimum
• X Scale Maximum
• Y Scale Minimum
• Y Scale Maximum
• Normalize trace data to a:
• Scale trace data to a new
carrier frequency of:
• Shift trace data DOWN by:
• Trace Smoothing Amount
• Power present at input of DUT
• AM Noise Measurement of an RF Signal
• AM Noise (dBc/Hz)
•10Hz
•100E+6Hz
•0dBc/Hz
• - 180 dBc/Hz
• 1 Hz bandwidth
• 1 times the current carrier frequency
•0dB
•0
•0dB
Table 13-3 Parameter Data for the AM Noise using an Agilent/HP 70420A Option
001
Step Parameters Data