Agilent Technologies E5500 Phase Noise Measurement System
AM Noise Measurement Examples
AM Noise using an Agilent/HP 70420A Option 001
NOTE The amplitude of a source under system control, for an AM noise
measurement, will automatically be set to +10 dBm. If any other amplitude
is desired, the source should be placed under manual control. All other
measurements set the source to +16 dBm automatically.
The appropriate measurement definition parameters for this example
have been pre-stored in this file. Table 13-3 on page 13-10 lists the
parameter data that has been entered for the FM Discriminator
measurement example.)
5. From the Define menu, choose Measurement; then choose the Type
and Range tab from the Define Measurement window.
a. From the Measurement Type pull-down, select AM Noise.