User`s guide
13-2
Agilent Technologies E5500 Phase Noise Measurement System
AM Noise Measurement Examples
AM Noise using an Agilent/HP 70420A Option 001
AM Noise using an Agilent/HP 70420A
Option 001
This example demonstrates the AM noise measurement of an
Agilent/HP 8662A Signal Generator using the AM detector in the
Agilent/HP 70420A Option 001 Phase Noise test set. For more information
about various calibration techniques, refer to Chapter 12, “AM Noise
Measurement Fundamentals”.
This measurement uses the double sided spur calibration method.
The measurement of a source with amplitude modulation capability is
among the simplest of the AM noise measurements. The modulation
sidebands used to calibrate the AM detector are generated by the DUT.
Required Equipment
CAUTION To prevent damage to the Agilent/HP 70420A test set’s hardware
components, the input signal must not be applied to the signal input
connector until the input attenuator has been correctly set for the desired
configuration, as show in Table 13-2 on page 13-7. Apply the input signal
when the Connection Diagram appears.
The following equipment is required for this example in addition to the
phase noise test system and your unit-under-test (UUT).
NOTE To ensure accurate measurements, you should allow the UUT and
measurement equipment to warm up at least one hour before making the
noise measurement.
The following is the configuration used for an AM noise measurement.
Table 13-1 Required Equipment for the AM Noise
using the Agilent/HP 70420A Option
001 Measurement Example
Equipment Quantity Comments
Agilent/HP 8644B 1
Coax Cables And adequate adapters to connect
the UUT and reference source to
the test set.