User`s guide

Agilent Technologies E5500 Phase Noise Measurement System
11-31
FM Discriminator Measurement Examples
Discriminator Measurement using FM Rate and Deviation Calibration
Table 11-6 Parameter Data for the Rate and Deviation Calibration Example
Step Parameters Data
1 Type and Range Tab
Measurement Type
Start Frequency
Stop Frequency
Minimum Number of Averages
FFT Quality
Swept Quality
Absolute Phase Noise (using an FM Discriminator)
•10Hz
•100E+6Hz
•4
•Normal
•Fast
2 Sources Tab
Carrier Source
Frequency
Power
Carrier Source is Connected to:
Detector Input
Frequency
1.027 E + 9 Hz
•19dBm
Test Set
1.027 E + 9 Hz
3CalTab
FM Discriminator Constant
Current Phase Detector
Constant
Know Spur Parameters
Offset Frequency
Amplitude
Calibration Source
Frequency
Power
Frequency Modulation
•FMRate
FM Deviation
Derive Constant from FM rate and deviation
82.25 E-9
•1E3
•-6dBc
1.027 E + 9 Hz
•16dBm
•20E+3Hz
•10E+3Hz
4 Block Diagram Tab
Carrier Source
Phase Shifter
•DUTinPath
Phase Detector
Adjust the Quadrature by
adjusting the
Manual
Manual
checked
Automatic Detector Selection
phase shifter