User`s guide
Agilent Technologies E5500 Phase Noise Measurement System
11-25
FM Discriminator Measurement Examples
Discriminator Measurement using FM Rate and Deviation Calibration
Beginning the
Measurement
1. From the View menu, choose Meter to select the quadrature meter.
2. From the Measurement menu, choose New Measurement
.
3. When the Perform a New Calibration and Measurement dialog box
appears, click OK.
4. When the Connect Diagram dialog box appears, confirm your
connections as shown in the connect kiagram. The Agilent/HP 70420A
test set’s signal input is subject to the following limits and
characteristics: