User`s guide
Agilent Technologies E5500 Phase Noise Measurement System
9-15
Residual Measurement Examples
Amplifier Measurement Example
4 Block Diagram Tab
• Carrier Source
• Phase Shifter
•DUTinPath
• Phase Detector
• Adjust the Quadrature by
adjusting the
• Manual
• Manual
• checked
• Automatic Detector Selection
• phase shifter
5 Test Set Tab
Input Attenuation
LNA Low Pass Filter
•LNAGain
•DCBlock
• PLL Integrator Attenuation
•0dB
• 20 MHz (Auto checked)
• Auto Gain (Minimum Auto Gain - 14 dB)
• Not checked
•0dBm
6 Dowconverter Tab • The downconverter parameters do not apply to this
measurement example.
7GraphTab
•Title
• Graph Type
• X Scale Minimum
• X Scale Maximum
• Y Scale Minimum
• Y Scale Maximum
• Normalize trace data to a:
• Scale trace data to a new
carrier frequency of:
• Shift trace data DOWN by:
• Trace Smoothing Amount
• Power present at input of DUT
• HP E5500 Residual Phase Noise Measurement @ 1 GHz.
• Single-sideband Noise (dBc/Hz)
•10Hz
•100E+6Hz
•0dBc/Hz
• - 180 dBc/Hz
• 1 Hz bandwidth
• 1 times the current carrier frequency
•0dB
•0
•0dB
Table 9-4 Parameter Data for the Amplifier Measurement Example
Step Parameters Data