User`s guide

Agilent Technologies E5500 Phase Noise Measurement System
8-21
Residual Measurement Fundamentals
The Calibration Options
Double-Sided Spur This calibration option has the following requirements:
One of the input frequency sources must be capable of being phase
modulated.
The resultant sideband spurs from the phase modulation must have
amplitudes that are –100 dB and –20 dB relative to the carrier
amplitude.
The offset frequency or modulation frequency must be between 10 Hz
and maximum (See the “Measured Beatnote” technique on page 8-16).
Advantages:
Requires only one RF source
Calibration is done under actual measurement conditions so all
non-linearities and harmonics of the phase detector are calibrated out.
NOTE Because the calibration is performed under actual measurement conditions,
the Double-sided Spur Method and the Single-sided Spur Method are the
two most accurate calibration methods.
Disadvantages:
Requires a phase modulator which operates at the desired carrier
frequency.
Requires audio calibration source.
Requires RF spectrum analyzer for manual measurement of ΦM
sidebands or preferably a modulation analyzer.
NOTE Most phase modulators are typically narrow-band devices; therefore, a wide
range of test frequencies may require multiple phase modulators.