User guide
48
Fragment Library Preparation Using the AB Library Builder
™
System: 5500 Series SOLiD
™
Systems User Guide
AppendixB Supplemental Procedures
AB Library Builder
™
System operation
B
IMPORTANT! Do not press against the cap to load or unload microTUBE vials,
because pressing against the cap may dislodge or damage the cap.
4. Release the plunger. The plunger pushes the tube until the base of the cap rests
against the prongs. The tube and holder are now ready to be inserted into the S
Series instrument.
Unload Covaris
®
microTUBE vials
1. Use a thumb to push the stainless steel plunger up into the body of the
microTUBE holder to relieve pressure on the cap.
2. Press against the side of the glass tube (not against the cap) to free the microTUBE
from the grip of the holder.
AB Library Builder
™
System operation
Use the front panel
Parts of the front panel
The front panel provides tools for operating the instrument and tools for the service
engineer to maintain the instrument: