Specifications
6
The operator can place the LNA in a “fixed” gain
configuration and select the amount of gain to
use (14 dB to 56 dB of gain in 14-dB steps), or the
operator can select “pause after auto LNA gain”
and manually adjust the LNA gain before the meas-
urement is started. It is also acceptable to manually
select an internal low-pass filter (LPF) to help
reduce the phase transients—being coincident with
the pulse, they will appear as PRF feedthrough
(which usually necessitates the use of a PRF filter).
When monitoring the phase transients of the base-
band signal (with a wide-band oscilloscope con-
nected to the monitor port of the 70420A), the
signal at the monitor port is 14 dB (5 times) lower
than the baseband signal.
PRF feedthrough
If the PRF feedthrough overloads the LNA for more
than 10 percent of the pulse width, measurement
accuracy can be degraded. If the magnitude of the
PRF feedthrough causes the LNA to be set to a
lower-than-appropriate gain or the baseband signal
analyzer to be set to a less sensitive range than
needed, overall system noise floor can be degraded.
The 70420A phase noise test set provides a group
of internal low-pass filters that can reduce the
magnitude of the PRF feedthrough signals. When
measuring pulsed carriers, it is recommended that
the operator place the test set LNA low-pass filters
in manual and select the filter that provides just
enough offset range for the pulsed measurement
(PRF/2). Refer to Table 2-1 for the proper selection
of the LNA low-pass filter. If these internal low-
pass filters are not sufficient for a particular PRF,
an external phase detector with external low-pass
filtering will be required.
Table 2-1. Selecting an Internal Low-Pass Filter
PRF LNA Low-Pass Filter
≤40 kHz 20 kHz
>40 kHz to ≤400 kHz 200 kHz
>400 kHz to ≤4 MHz 2 MHz
>4 MHz to ≤40 MHz 20 MHz
Figure 2-2. Phase transients (100 MHz LPF & max LNA gain) as observed on
a scope connected to the monitor port of the Agilent 70420