b) When the connect diagram is displayed,
check hardware and connections.
1) Connect scope to monitor port of 70420A
test set.
2) Set scope to monitor waveform.
3) Connect the signal paths to the phase noise
test set.
4) Set quadrature (set the signal to zero volts
on scope and a null on View Meter).
c) Select Continue.
d) When prompted by the software, connect the
calibration signal from the calibration source
to the coupler, then select Continue.
e) When prompted by the software, disconnect
the calibration signal from the calibration
source, then select Continue.
8. When the measurement is complete, print the
results and save the measurement file.
9. Interpret the results. Note the PRF feedthrough,
aliasing effects, and any breaks in the graph. If
PRF feedthrough is high and there are breaks in
the graph, set the LNA Gain to 56 dB of gain and
watch for clipping on the monitor oscilloscope.
Start a Repeat measurement and adjust the quad-
rature at the connect diagram.
Figure 2-14. Pulsed carrier amplifier residual measurement with 200-kHz LPF