User`s guide

Table Of Contents
Chapter 13 485
SCPI Command Reference
E5061A/E5062A commands
13. SCPI Command
Reference
:TRIG:SOUR
Syntax :TRIGger[:SEQuence]:SOURce {INTernal|EXTernal|MANual|BUS}
:TRIGger[:SEQuence]:SOURce?
Description Selects the trigger source from the following 4 types.
Internal Uses the internal trigger to generate continuous triggers automatically.
External Generates a trigger when the trigger signal is inputted externally via
the Ext Trig connector or the handler interface.
Manual Generates a trigger when the key operation of
[Trigger] - Trigger is
executed from the front panel.
Bus Generates a trigger when the *TRG command is executed.
When you change the trigger source during sweep, the sweep is canceled.
Parameters
Query response {BUS|EXT|INT|MAN}<newline><^END>
Example of use
10 OUTPUT 717;":TRIG:SOUR BUS"
20 OUTPUT 717;":TRIG:SOUR?"
30 ENTER 717;A$
Related commands *TRG on page 201
Equivalent key
[Trigger] - Trigger Source - Internal|External|Manual|Bus
Description
INTernal (preset value) Specifies internal.
EXTernal Specifies external.
MANual Specifies manual.
BUS Specifies bus.